![Rohde & Schwarz 1145.1010.04/05/06 Скачать руководство пользователя страница 247](http://html1.mh-extra.com/html/rohde-and-schwarz/1145-1010-04-05-06/1145-1010-04-05-06_operating-manual_1477759247.webp)
R&S
®
ZVA/ZVB/ZVT
Trace
Menu
1145.1084.12 4.85
E-6
•
The
Real
added constant shifts a real trace in vertical direction, leaving the imaginary part
unchanged.
•
The
Imaginary
added constant shifts a imaginary trace in vertical direction, leaving the real part
unchanged.
Shifting the trace by means of constant values is a simple case of trace mathematics. Use the
Define Math
dialog to define more complicated mathematical operations.
Remote
control:
DISPlay:WINDow<Wnd>:TRACe<WndTr>:Y:OFFset <Magnitude>[, <Phase>,
<Real>, <Imag>]
Marker
The functions in the
Marker
menu are used to position markers on a trace, configure their properties
and select the format of the numerical readout.
Markers
Markers are tools for selecting points on the trace and for numerical readout of measured data. A
marker is displayed with a symbol (a triangle, a crossbar or a line) on the trace, which may be a data
trace or a memory trace. At the same time, the coordinates are displayed in a marker info field or in a
table. Each marker can be defined as a normal marker, reference marker, delta marker or discrete
marker.
Marker types
•
A (normal) marker
(Mkr 1, Mkr 2, ...)
determines the coordinates of a measurement point on the
trace. Up to 10 different normal markers can be assigned to a trace.
•
A reference marker
(Ref)
defines the reference value for all delta markers.
•
A delta marker
( )
indicates the coordinates relative to the reference marker.
•
The stimulus value of a
discrete marker
always coincides with a sweep point.
A special set of markers Mk1 to Mkr4 is provided for bandfilter search mode.
The most common tasks to be performed with markers can be achieved using the
Marker
menu
functions:
•
Determine the coordinates of a measurement point on the trace. In polar diagrams where no x-
axis is displayed, markers can be used to retrieve the stimulus value of specific points.
•
Determine the difference between two trace points or the relative measurement result (
Delta
Mode
).
•
Convert a complex measurement result into other formats.
Markers also play an important role in performing the following advanced tasks:
•
Change the sweep range and the diagram scale (
Marker Funct.
).
•
Search for specific points on the trace (
Search
).
Содержание 1145.1010.04/05/06
Страница 10: ......
Страница 20: ......
Страница 22: ......
Страница 24: ...Contents of Chapter 1 R S ZVA 1145 1090 62 I 1 2 E 1 Windows XP Embedded 1 22 Firmware Update 1 22...
Страница 48: ......
Страница 70: ......
Страница 72: ......
Страница 90: ......
Страница 92: ......
Страница 108: ......
Страница 156: ......
Страница 162: ......
Страница 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Страница 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Страница 462: ......
Страница 766: ......
Страница 772: ......
Страница 792: ......
Страница 794: ......
Страница 808: ......