R&S
®
ZVA/ZVB/ZVT
Channel
Menu
1145.1084.12 4.199
E-6
One Port P1, One Port P2
Opens a submenu to select a one-port calibration at test ports PORT 1 or PORT 2. The two submenus
are identical:
One-port calibrations can be performed automatically or manually.
•
Calibration Unit
starts an automatic full one-port calibration. This function is enabled while the
Calibration Unit is connected.
The following menu commands calls up the start dialog of the calibration wizard to start a manual
calibration:
•
Normalization (Short)
initiates a normalization using a short standard.
•
Full
initiates a full one-port calibration.
Remote
control:
[SENSe<Ch>:]CORRection:COLLect:AUTO ' ', <port_no>
(for matching
analyzer and cal unit ports)
[SENSe<Ch>:]CORRection:COLLect:AUTO:PORTs
(for arbitrary pairs of ports)
[SENSe<Ch>:]CORRection:COLLect:METHod:DEFine
"<cal_name>", REFLshort | FOPort , <port_no>
Two Port P1 P2
Opens a submenu to select a two-port calibration at test ports PORT 1 and PORT 2:
Содержание 1145.1010.04/05/06
Страница 10: ......
Страница 20: ......
Страница 22: ......
Страница 24: ...Contents of Chapter 1 R S ZVA 1145 1090 62 I 1 2 E 1 Windows XP Embedded 1 22 Firmware Update 1 22...
Страница 48: ......
Страница 70: ......
Страница 72: ......
Страница 90: ......
Страница 92: ......
Страница 108: ......
Страница 156: ......
Страница 162: ......
Страница 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Страница 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Страница 462: ......
Страница 766: ......
Страница 772: ......
Страница 792: ......
Страница 794: ......
Страница 808: ......