SENSe
R&S
®
ZVA/ZVB/ZVT
1145.1084.12 6.202
E-6
[SENSe<Ch>:]CORRection:COLLect:METHod:DEFine '<cal_name>', REFL |
RSHort | FOPort | FRTRans | OPTPort | TOSM | TOM | TRM | TRL | TNA | ETOM |
ETSM, <port_no>[,<port_no>][,<port_no>][,<port_no>]
Selects a one-port, two-port, three-port or four-port calibration type at arbitrary analyzer ports.
<Ch>
Channel number of the calibrated channel.
'<cal_name>'
Name of the calibration (string parameter). The name serves as a
reference to delete a particular set of system correction data
(
[SENSe<Ch>:]CORRection:COLLect:DELete
).
Parameters
Calibration types:
One-port Normalization (reflection), using an open
(REFL)
or a short (RSHort) standard, Full One Port, Two-port
Normalization (transmission,
FRTRans)
, One Path Two Port, TOSM
,
TOM, TRM, TRL, TNA, TOM Enhanced, TSM Enhanced
.
<port_no>
Port numbers of the analyzer. For an n-port calibration type (n = 1 to
4), n port numbers must be specified. If more than n numbers are
defined, the spare numbers (the last ones in the list) are ignored.
Entering less than n numbers causes an error message.
*RST value
–
SCPI,
Command
Types
Device-specific, no query. Use
[SENSe<Ch>:]CORRection:COLLect:METHod?
to obtain a list of
all calibration types for channel <Ch>.
Example:
See [SENSe<Ch>:]
CORRection:COLLect:SAVE:SELected
.
[SENSe<Ch>:]CORRection:COLLect:SAVE
Calculates the system error correction data from the acquired one or two-port measurement results
(
[SENSe<Ch>:]CORRection:COLLect[:ACQuire]),
stores them and applies them to the
calibrated channel <Ch>. To avoid incompatibilities, older system error correction data is deleted unless
it has been transferred into a cal pool (
MMEMory:STORe:CORRection <Ch>, '<file_name>')
.
This command is the ZVR-compatible equivalent of
[SENSe<Ch>:]CORRection:COLLect:SAVE:SELected
.
It must be used in combination with
the ZVR-compatible commands for calibration method and standard selection; see
example below.
<Ch>
Channel number of the calibrated channel.
*RST value
–
SCPI, Command
Types
Confirmed, no query
Содержание 1145.1010.04/05/06
Страница 10: ......
Страница 20: ......
Страница 22: ......
Страница 24: ...Contents of Chapter 1 R S ZVA 1145 1090 62 I 1 2 E 1 Windows XP Embedded 1 22 Firmware Update 1 22...
Страница 48: ......
Страница 70: ......
Страница 72: ......
Страница 90: ......
Страница 92: ......
Страница 108: ......
Страница 156: ......
Страница 162: ......
Страница 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Страница 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Страница 462: ......
Страница 766: ......
Страница 772: ......
Страница 792: ......
Страница 794: ......
Страница 808: ......