Trace Evaluation
R&S
®
ZVA/ZVB/ZVT
1145.1084.12 9.8
E-4
Trace Evaluation
Time Domain Measurements
With option ZVAB-K2,
Time Domain,
you can view the measurement results as a function of time.
To set up a time domain measurement...
1. Reset the analyzer to ensure that the following operating mode is set:
Sweep Type: Lin.
Frequency, Trace - Meas.: S21, Trace - Format: dB Mag
.
2. Open
the
Trace - Meas
menu and select the measured quantity to be transformed (preferably:
an S-parameter or ratio).
3. Click
Trace - Trace Func - Transform - Time Domain
to transform the active trace to the time
domain.
The transformed trace is displayed in the active diagram area. To refine the analysis, you can modify
the number and position of the sweep points
(Channel - Sweep - Number of Points, Channel -
Stimulus),
select a different transform type and window for profiling
(Trace - Trace Func - Transform -
Define Transform...)
and vary the time range
(Trace - Trace Func - Transform - Time Domain Stimulus
Axis...).
To eliminate an unwanted response from the measured trace...
1. Calculate a time domain transform as descibed above, Setting up a time domain measurement.
2. Click
Trace - Trace Func - Transform - Define Time Gate
and
place a time gate on the
transformed trace. To remove a single response use a
Notch
gate type. To everything outside a
specified area, use a
Bandpass
gate type.
3. Click
Trace - Trace Func - Transform - Time Gate
to enable the time gate. If necessary, go
back to the
Define Time Gate
dialog to correct your time gate settings.
4. Click
Trace - Trace Func - Transform - Frequency Domain.
The frequency domain trace now corresponds to the modified time domain trace which is transformed
back into the frequency domain
.
Trace Statistics
After measuring a quantity across the selected sweep range, the analyzer can analyze the trace and
determine various statistical results.
To determine the x-dB compression point...
1. Press
the
Sweep
key in the
Channel
keypad or click
Channel - Sweep
to access the Sweep
submenu.
2. Click
Sweep Type - Power.
In the
CW Frequency
input field opened, enter the fixed stimulus
frequency for the power sweep.
3. Click
Channel - Stimulus - Start
and
Channel - Stimulus - Start
to define the power sweep
range.
Содержание 1145.1010.04/05/06
Страница 10: ......
Страница 20: ......
Страница 22: ......
Страница 24: ...Contents of Chapter 1 R S ZVA 1145 1090 62 I 1 2 E 1 Windows XP Embedded 1 22 Firmware Update 1 22...
Страница 48: ......
Страница 70: ......
Страница 72: ......
Страница 90: ......
Страница 92: ......
Страница 108: ......
Страница 156: ......
Страница 162: ......
Страница 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Страница 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Страница 462: ......
Страница 766: ......
Страница 772: ......
Страница 792: ......
Страница 794: ......
Страница 808: ......