SENSe
R&S
®
ZVA/ZVB/ZVT
1145.1084.12 6.216
E-6
[SENSe<Ch>:]FREQuency:CONVersion FUNDamental | HARMonic | SHARmonic
| THARmonic | MIXer
Enables the measurement of a harmonic or the fundamental signal for channel <Ch>.
Enabling a harmonics measurement also resets the measured quantity; see program example
below.
<Ch>
Channel number.
FUNDamental
Measurement of the fundamental signal (no frequency-converting
mode).
HARMonic
Harmonics measurement. The order of the measured harmonic, the
source port, and the receive port must be selected using the
[SENSe<Ch>:]FREQuency:CONVersion:HARMonic...
commands; see example below.
SHARmonic |
THARmonic
Measurement of the 2nd or 3rd harmonic with source port 1 and
receive port 2. These parameters have been introduced for
compatibilty with the ZVR. While they are active, a query returns
HARM
.
MIXer
Mixer mode. The mixer mode is configured using the
[SENSe<Ch>:]FREQuency:CONVersion:MIXer...
commands.
*RST value
FUNDamental
SCPI,
Command
Types
Device-specific, command or query. Returned values:
FUND
or
HARM
.
Example:
*RST; FREQ:STAR 0.5E+9; STOP 1E+9
Reset the analyzer (activating a frequency sweep) and set the sweep
range between 1 GHz and 1.1 GHz.
FREQ:CONV:HARM:ORD 4; SPOR 1; RPOR 2
Configure a 4
th
harmonic measurement with source port 1 and receive
port 2. The measurement is still not enabled.
FREQ:CONV HARM
Enable the configured harmonic measurement.
CALC:PAR:MEAS? "TRC1"
Query the measured quantity. The analyzer measures the wave
quantity b2, the response is
'B2D1SAM'
.
FREQ:CONV:HARM:RELative ON
Enable the relative harmonic measurement, creating a new channel
and a new trace
Trc2
.
CALC2:PAR:MEAS? "TRC2"
Query the measured quantity in the new channel no. 2. The analyzer
still measures the wave quantity b2, the response is
'B2D1SAM'
.
CALC2:MATH:STAT?
Check whether the active trace in channel no. 2 (i.e.
Trc2)
is a
mathematical trace
:
the response is
1
.
Содержание 1145.1010.04/05/06
Страница 10: ......
Страница 20: ......
Страница 22: ......
Страница 24: ...Contents of Chapter 1 R S ZVA 1145 1090 62 I 1 2 E 1 Windows XP Embedded 1 22 Firmware Update 1 22...
Страница 48: ......
Страница 70: ......
Страница 72: ......
Страница 90: ......
Страница 92: ......
Страница 108: ......
Страница 156: ......
Страница 162: ......
Страница 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Страница 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Страница 462: ......
Страница 766: ......
Страница 772: ......
Страница 792: ......
Страница 794: ......
Страница 808: ......