Channel Menu
R&S
®
ZVA/ZVB/ZVT
1145.1084.12 4.134
E-6
Sweep Type
The
Sweep Type
submenu defines the sweep variable (frequency/power/time) and the position of the
sweep points across the sweep range.
•
Lin. Frequency
is the default sweep type. The stimulus frequency is swept in equidistant steps
over the continuous frequency range. In a Cartesian diagram, the x-axis is a linear frequency
axis.
•
Log. Frequency
is
analogous to
Lin. Frequency,
however, the frequency is swept in equidistant
steps on a logarithmic scale. In a Cartesian diagram, the x-axis is a logarithmic frequency axis.
•
Segmented Frequency
is
analogous to
Lin. Frequency
but uses a sweep range that can be
composed of several continuous frequency ranges or single frequency points defined via Define
Segments.
A valid segment list must be defined before activating
Segment Frequency.
•
Power
sweeps are performed at constant frequency but with variable generator power that is
swept in linear, equidistant steps over a continuous range. In a Cartesian diagram, the x-axis is
a dB-linear power axis.
•
Time
sweeps are performed at constant frequency and stimulus power. The measurement is
repeated over a specified period of time at constant time intervals.
•
CW Mode
is a variant of the
Time
sweep where the individual measurements are triggered
according to the current trigger settings.
Lin. Frequency
In a
Lin. Frequency
sweep the stimulus frequency is swept in equidistant steps over the continuous
frequency range. The frequency range (sweep range) is defined with the Stimulus settings. The step
width between two consecutive sweep points is constant and given by <Span>/(n - 1) where n is the
specified Number of Points (n > 1). The internal generator power can be set, if so desired, in the Power
Bandwidth Average submenu.
A
Lin. Frequency
sweep corresponds to the analysis of a signal over the frequency, as obtained e.g. by
means of a spectrum analyzer. This is the default sweep type. In a Cartesian diagram the measurement
result is displayed as a trace over a linear frequency scale (spectral representation). The following
example shows a
Lin. Frequency
sweep with a stimulus range between 4.5 GHz and 6 GHz, the
forward transmission parameter S12 as measured quantity, and a
dB Mag
scaled y-axis.
Содержание 1145.1010.04/05/06
Страница 10: ......
Страница 20: ......
Страница 22: ......
Страница 24: ...Contents of Chapter 1 R S ZVA 1145 1090 62 I 1 2 E 1 Windows XP Embedded 1 22 Firmware Update 1 22...
Страница 48: ......
Страница 70: ......
Страница 72: ......
Страница 90: ......
Страница 92: ......
Страница 108: ......
Страница 156: ......
Страница 162: ......
Страница 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Страница 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Страница 462: ......
Страница 766: ......
Страница 772: ......
Страница 792: ......
Страница 794: ......
Страница 808: ......