Channel Menu
R&S
®
ZVA/ZVB/ZVT
1145.1084.12 4.118
E-6
Channel Menu
The
Channel
menu provides all channel settings and the functions to activate, modify and store different
channels.
Channels
A channel contains hardware-related settings to specify how the network analyzer collects data. The
channel settings can be divided into three main groups:
•
Control of the measurement process
(Sweep, Trigger, Average)
•
Description of the test setup
(internal source power, IF filters and step attenuators)
•
Correction data
(Calibration, Offset)
The channel settings complement the definitions of the
Trace
menu. Each trace is assigned to a
channel, see Traces, Channels and Diagram Areas. The channel settings apply to all traces assigned
to the channel.
The
Channel
menu contains the following functions and submenus:
•
Stimulus
defines the sweep range, depending on the sweep type.
•
Power Bandwidth Average
defines the power of the internal signal source, sets the step
attenuators and the IF bandwidths, and configures the sweep average.
•
Sweep
defines the scope of measurement, including the sweep type, the trigger conditions and
the periodicity of the measurement.
•
Mode
opens the
Port Configuration
dialog to define the properties of the physical and logical
(balanced) test ports.
•
Channel Select
provides functions to handle and activate channels.
•
Calibration
provides all functions that are necessary to perform a system error correction
(calibration).
•
Offset
provides a selection of length offset parameters to shift the measurement plane.
Stimulus
The
Stimulus
submenu defines the sweep range in the current channel, depending on the sweep type.
•
In Cartesian diagrams, the sweep range corresponds to the diagram width and defines the
scaling of the x-axis.
•
In polar diagrams and Smith charts the stimulus axis is lost but the marker functions easily
provide the stimulus value of any measurement point.
Содержание 1145.1010.04/05/06
Страница 10: ......
Страница 20: ......
Страница 22: ......
Страница 24: ...Contents of Chapter 1 R S ZVA 1145 1090 62 I 1 2 E 1 Windows XP Embedded 1 22 Firmware Update 1 22...
Страница 48: ......
Страница 70: ......
Страница 72: ......
Страница 90: ......
Страница 92: ......
Страница 108: ......
Страница 156: ......
Страница 162: ......
Страница 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Страница 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Страница 462: ......
Страница 766: ......
Страница 772: ......
Страница 792: ......
Страница 794: ......
Страница 808: ......