Series 8650B Universal Power Meters
3-116
Publication 31470-001, Rev C, November 2, 2017
3.18
Measurement Mode Commands
3.18.1
CW Mode
Commands which specify the CW measurement mode are based on the CW function code:
Syntax:
CW [A or B]
[AE or BE] specifies Sensor A or Sensor B
These commands can be used with any sensor (although it is superfluous in the case of a CW sensor). Possible
remote entry errors: 60/61 (uncalibrated or missing sensor A/B). Refer to Table 3-22 for error numbers.
Example:
CW A
! SELECTS CW MODE FOR SENSOR A
Example 2:
CW B
! SELECTS CW MODE FOR SENSOR B
3.18.2
MAP Mode
Commands which specify the Modulated Average Power measurement mode are based on the MAP function
code:
Syntax:
MAP [A or B]
[AE or BE] specifies Sensor A or Sensor B
These commands will work only with a modulated sensor. Possible remote entry errors: 60/61 (uncalibrated or
missing sensor A/B), 62/63 (not a modulated sensor, or two sensor operation active, A/B). Refer to Table 3-22
for error numbers.
Example:
MAP A
! SELECTS MAP MODE FOR SENSOR A
Example 2:
MAP B
! SELECTS MAP MODE FOR SENSOR B
If an irregularly modulated signal is measured in MAP mode, measurement settling time will vary as the power
meter attempts to synchronize to the modulation. In such a situation, it may be desirable to disable
synchronization for faster measurement. The commands which disable synchronization are based on the MAP
function code.
Syntax:
MAP [A or B] 0
[AE or BE] specifies Sensor A or Sensor B
0 specifies that synchronization is to be disabled
Example:
MAP A 0
! DISABLES MAP MODE SYNCHRONIZATION FOR SENSOR A
Example 2:
MAP B 0
! DISABLES MAP MODE SYNCHRONIZATION FOR SENSOR B
NOTE:
To reactivate synchronization, send the MAP A or MAP B command again.
Содержание 8650B Series
Страница 3: ...Series 8650B Universal Power Meters Publication 31470 001 Rev C November 2 2017 iii ...
Страница 18: ......
Страница 32: ......
Страница 78: ......
Страница 236: ......
Страница 244: ......
Страница 258: ......
Страница 263: ...Menu Structure Publication 31470 001 Rev C November 2 2017 D 3 Figure D 3 Modulation Sensor B Setup Menu Structure ...
Страница 265: ...Menu Stucture Publication 31470 001 Rev C November 2 2017 Page D 5 Figure D 4 Meter Setup Menu Structure ...
Страница 267: ...Publication 31470 001 Rev C November 2 2017 Last page of the document ...