Performance Verification
Publication 31470-001, Rev C, November 2, 2017
4-1
Chapter 4
Performance Verification
4.1
Introduction
This chapter defines procedures for performance verification of the Series 8650B Universal Power Meters. Refer
to Appendix B of this publication for power sensor selection and calibration.
4.2
Equipment Required
Table 4-1 lists the equipment required for performance verification of the Series 8650B Power Meters.
Table 4-1: Performance Verification – Required Equipment List
Description
Representative Model
Key Characteristics
CW Thermistor Power Meter
HP/Agilent
Model 432A
V
RF
and V
COMP
available externally
Thermistor Mount
HP/Agilent
Model 478A-H75
≤1.05 VSWR @ 50 MHz
Accuracy ±0.5% @ 50 MHz
Digital Voltmeter (DVM)
HP/Agilent
Model 34401A
±0.05% accuracy
1 µV resolution
Directional Coupler, 10 dB
Mini-Circuits
Model ZFDC-10-1-N+
≤1.20 SWR @ 50 MHz
Step Attenuator, 0 to 90 dB in
10 dB increments
Aeroflex/Weinschel
Model AC118A-90-43
≤1.15 SWR @ 50 MHz
±0.1 dB attenuation
RF Source (Signal Generator)
(RF Amp optional)
Giga-tronics
Model GT-2500B
+22 dBm @ 50 MHz
Low Pass Filter, 50 MHz
Mini-Circuits
Model NLP-50+
>30 dB attenuation @ 100 MHz
Frequency Counter
XL Microwave
Model 3260
≥1.5 GHz Range
≥6 digit resolution
≤50 ppm stability
GPIB Controller
PC w/ GPIB interface
With driver software
Содержание 8650B Series
Страница 3: ...Series 8650B Universal Power Meters Publication 31470 001 Rev C November 2 2017 iii ...
Страница 18: ......
Страница 32: ......
Страница 78: ......
Страница 236: ......
Страница 244: ......
Страница 258: ......
Страница 263: ...Menu Structure Publication 31470 001 Rev C November 2 2017 D 3 Figure D 3 Modulation Sensor B Setup Menu Structure ...
Страница 265: ...Menu Stucture Publication 31470 001 Rev C November 2 2017 Page D 5 Figure D 4 Meter Setup Menu Structure ...
Страница 267: ...Publication 31470 001 Rev C November 2 2017 Last page of the document ...