Remote Operation
Publication 31470-001, Rev C, November 2, 2017
3-39
3.2.19
Peak Power Sensor Triggering
SENSe:TRIGger
The Peak Power Sensors sample power levels almost instantaneously. Since there is a sampler built into the Peak
Power Sensor housing, there are several controls to configure the source of the sensor trigger signal. These
include the delay time from triggering to when the sample is to be taken (Sample Delay), and the trigger level.
All Peak Power Sensors will operate in the GPIB SWIFt and GPIB BURSt modes when in the CW measurement
mode configuration.
SENSe:TRIGger
Syntax:
SENSe<sensor 1 or 2>:TRIGger:SOURcespace<INTernal, EXTernal, or CW>
Example:
SENS1:TRIG:SOUR CW
! SETS PEAK SENSOR 1 TRIGGER SOURCE TO CW MODE
Description:
This command controls the trigger source for the 80350A Series Peak sensors.
INTernal triggering is performed automatically by the 80350A Sensor; however, the trigger
level must be set to a value of power that is lower than the power level of the pulsed signal
incident upon the sensor.
EXTernal triggering is performed by inputting a baseband pulse into the rear connector on the
sensor housing.
CW triggering sets the power sensors to measure CW power and triggering is essentially
disabled. CW is automatically set by the 8650B during power sweep calibration; do not apply
any triggering to the peak power sensor during power sweep calibration.
Be sure to set the trigger level, SENS:TRIG:LEV ##.# after the INTernal or EXTernal triggering
is selected.
Syntax:
SENSe<sensor 1 or 2>:TRIGger:DELay[:MAGnitude]space<delay in seconds from -50.5e-9 to 104.8e-3>
Example:
SENS1:TRIG:DEL 1E-6
! SETS PEAK SENSOR 1 DELAY VALUE TO 1E-6 SECONDS
Description:
This command controls the delay time in seconds between the 80350A Series Peak Power
Sensor’s receipt of a sensor trigger (INT is auto, EXT is a trigger input) and the time that the
sampler is fired. This delay is typically accurate to about 0.5%; repeatability is an order of
magnitude better. There is a delay line built into the peak power sensor which allows limited
look-ahead capability as identified by the allowed negative time values for this command. The
reason negative time values are allowed is due to the sample delay offset capability. Therefore,
measuring the power on the leading edge of the pulsed signal is possible, even if it is the same
edge on which the sensor was triggered.
Syntax:
SENSe<sensor 1 or 2>:TRIGger:DELay:STATespace<ON | OFF>
Example:
SENS1:TRIG:DEL:STAT ON
! ENABLES PEAK SENSOR 1 DELAY FUNCTION
Description:
This command enables or disables the sample delay in the 80350A Series Peak Power
Sensors.
Содержание 8650B Series
Страница 3: ...Series 8650B Universal Power Meters Publication 31470 001 Rev C November 2 2017 iii ...
Страница 18: ......
Страница 32: ......
Страница 78: ......
Страница 236: ......
Страница 244: ......
Страница 258: ......
Страница 263: ...Menu Structure Publication 31470 001 Rev C November 2 2017 D 3 Figure D 3 Modulation Sensor B Setup Menu Structure ...
Страница 265: ...Menu Stucture Publication 31470 001 Rev C November 2 2017 Page D 5 Figure D 4 Meter Setup Menu Structure ...
Страница 267: ...Publication 31470 001 Rev C November 2 2017 Last page of the document ...