Remote Operation
Publication 31470-001, Rev C, November 2, 2017
3-63
3.2.33
Calibrator Controls
OUTPut:Reference:OSCillator
OUTPut:ReferenceOSCillator
Syntax:
OUTPut:ReferenceOSCillator[:STATe]
space
<ON | OFF>
Example:
OUTP:ROSC ON
! ENABLES THE 8650B REFERENCE OSCILLATOR OUTPUT
Description:
The OUTPut:ReferenceOSCillator ON command turns on the reference oscillator 50 MHz, 0
dBm output. The reference oscillator (Calibrator Output port) connection is on the front panel of
the power meter. The reference oscillator is the RF calibration source for the sensors. The
reference oscillator power level should be calibrated annually.
3.2.34
Sensor EEPROM Commands
SENSe:CORRection:EEPROM:TYPE?
SENSe:CORRection:EEPROM:CALFactor?
SENSe:CORRection:EEPROM:FREQuency?
SENSe:CORRection:EEPROM:TYPE?
Syntax:
SENSe<sensor 1 or 2>:CORRection:EEPROM:TYPE?
Example:
SENS1:CORR:EEPROM:TYPE?
! QUERIES SENSOR 1 EEPROM TYPE
Response:
80301,1818436 (which is sensor model number and serial number respectively)
Description:
This command returns sensor model and serial number information. This command allows
programming to verify correct sensor usage and monitor whether power sweep calibration has
been performed or not.
SENSe:CORRection:EEPROM:FREQuency?
Syntax:
SENSe<sensor 1 or 2>:CORRection:EEPROM:FREQuency?
Example:
SENS1:CORR:EEPROM:FREQ?
! QUERIES SENSOR 1 EEPROM FREQUENCY TABLE
Response:
Typically: 5.0000e7, 2.0000e9, 3.0000e9, …
Description:
This command returns a listing of the calibration frequencies stored in the sensor EEPROM,
sorted by frequency, so that both standard and special frequencies are listed. The number of
items matches the sum of the number at standard frequencies, and the number of special
frequencies.
SENSe:CORRection:EEPROM:CALFactor?
Syntax:
SENSe<sensor 1 or 2>:CORRection:EEPROM:CALFactor?
Example:
SENS1:CORR:EEPROM:CALF?
! QUERIES SENSOR 1 EEPROM CAL FACTOR TABLE
Response:
Typically: 0.0000e0, -4.0000e-2, -6.0000e-2, …
Description:
This command returns a listing of the calibration factors stored in the sensor EEPROM, sorted in
frequency order, to match the previous command. The number of items matches the sum of the
number at standard frequencies, and the number of special frequencies reported by
DIAG:SENS1:EEPROM:CALFR?
Содержание 8650B Series
Страница 3: ...Series 8650B Universal Power Meters Publication 31470 001 Rev C November 2 2017 iii ...
Страница 18: ......
Страница 32: ......
Страница 78: ......
Страница 236: ......
Страница 244: ......
Страница 258: ......
Страница 263: ...Menu Structure Publication 31470 001 Rev C November 2 2017 D 3 Figure D 3 Modulation Sensor B Setup Menu Structure ...
Страница 265: ...Menu Stucture Publication 31470 001 Rev C November 2 2017 Page D 5 Figure D 4 Meter Setup Menu Structure ...
Страница 267: ...Publication 31470 001 Rev C November 2 2017 Last page of the document ...