Series 8650B Universal Power Meters
3-20
Publication 31470-001, Rev C, November 2, 2017
3.2.10.2
Sensor Zero
Zeroing automatically accounts for ground noise and other noise in the measurement system. Measurements
will be sensitive to noise-induced errors only in the lowest 15 dB of the sensor dynamic range.
Be sure to turn off
the signal going into the sensor during zeroing, otherwise a failure will be indicated
.
CALibrate:ZERO
Syntax:
CALibration<sensor 1 or 2>:ZERO
Example:
CAL1:ZERO
! ZERO SENSOR 1
Description:
This command begins the zeroing process. To offset for measurement circuit noise during
zeroing, The sensor needs to remain attached to the measurement circuit during zeroing.
Disable the signal source into the sensor. If disabling the signal source cannot occur, connect
the sensor to a grounded connector (preferably RF grounded) or leave the sensor
disconnected. Do not connect the sensor to the power meter front panel calibrator port. If the
meter detects excessive RF power at the start of the zeroing process (generally above -50
dBm for standard sensors), zeroing will automatically fail.
SENSe:TEMP?
Syntax:
SENSe<sensor 1 or 2>:TEMPerature?
Example:
SENS1:TEMP?
! QUERIES SENSOR 1 TEMPERATURE (IN CENTIGRADE)
Response:
Current sensor temperature in degrees centigrade
Description:
This command reports sensor temperature. If the temperature varies in the operating
environment, monitor the relative temperature variations. If the variation since the previous
power sweep calibration exceeds ±5 °C (±9 °F), perform the power sweep calibration for that
sensor.
NOTE:
Example programs Sensor Calibration and Zeroing are contained in Appendix A of this publication.
Содержание 8650B Series
Страница 3: ...Series 8650B Universal Power Meters Publication 31470 001 Rev C November 2 2017 iii ...
Страница 18: ......
Страница 32: ......
Страница 78: ......
Страница 236: ......
Страница 244: ......
Страница 258: ......
Страница 263: ...Menu Structure Publication 31470 001 Rev C November 2 2017 D 3 Figure D 3 Modulation Sensor B Setup Menu Structure ...
Страница 265: ...Menu Stucture Publication 31470 001 Rev C November 2 2017 Page D 5 Figure D 4 Meter Setup Menu Structure ...
Страница 267: ...Publication 31470 001 Rev C November 2 2017 Last page of the document ...