Remote Operation
Publication 31470-001, Rev C, November 2, 2017
3-97
3.11.2.6
HOLDOFF h
HOLDOFF
h
specifies the holdoff time from the end of the gating period to the time when the circuit will accept
another trigger input edge. This parameter is relevant only in the External Trigger Mode. Since the previous set
up may be unknown, the DELAY, DURATION and HOLDOFF should be sent in series. Be aware that this command
is set up such that it automatically enables External Trigger Mode. That is, if this command is sent, the External
Trigger Mode will be enabled on the specified channel along with the specified holdoff delay.
Syntax:
GATE [A or B] HOLDOFF [h]
GATE is the function code for the Timing Gate function.
[A or B] specifies Sensor A or Sensor B.
HOLDOFF [h] specifies the holdoff time from the end of the gating period to the time when the
circuit will accept another trigger input edge. Variable [h] must be a time specified in seconds in
the range of 0 to 100 ms, with a resolution of 1 μs.
Example:
GATE A HOLDOFF 80E-3
! SPECIFIES A 80 mS HOLDOFF TIME FOR THE SENSOR A TIME GATE
! ALSO ENABLES EXTERNAL TRIGGER MODE
Example 2:
GATE B DURATION 950E-6
! SPECIFIES A 950 µS HOLDOFF TIME FOR THE SENSOR B TIME GATE
! ALSO ENABLES EXTERNAL TRIGGER MODE
3.11.2.7
Error Codes
If specifying an out of range value for the delay, gate duration, or holdoff values, the Entry Error bit of the Status
Byte Register is also set as well as the Execution Error bit of the Event Status Register. If requesting the Status
Message, the code in the Entry Error portion will be set.
See Section 3.19.13.3 for more information.
Содержание 8650B Series
Страница 3: ...Series 8650B Universal Power Meters Publication 31470 001 Rev C November 2 2017 iii ...
Страница 18: ......
Страница 32: ......
Страница 78: ......
Страница 236: ......
Страница 244: ......
Страница 258: ......
Страница 263: ...Menu Structure Publication 31470 001 Rev C November 2 2017 D 3 Figure D 3 Modulation Sensor B Setup Menu Structure ...
Страница 265: ...Menu Stucture Publication 31470 001 Rev C November 2 2017 Page D 5 Figure D 4 Meter Setup Menu Structure ...
Страница 267: ...Publication 31470 001 Rev C November 2 2017 Last page of the document ...