Series 8650B Universal Power Meters
3-84
Publication 31470-001, Rev C, November 2, 2017
3.6
Cal Factors
The commands described below do not need to be employed with the 8650B; it is included here for the sake of
compatibility with remote programs written for older power meters.
When a sensor is attached to the 8650B, the power meter automatically loads calibration factors from an
EEPROM in the sensor. This data is frequency related, and in order for the 8650B to make use of it, the user
must supply frequency information to the power meter, either by means of the front panel FREQ key, by means
of the GPIB FR command (See FREQUENCY, Section 3.12.2), or by means of the V
PROP
F input. Once the frequency
has been specified, the 8650B automatically applies the appropriate cal factor to each reading.
The KB function code specifies a cal factor which is to be used in place of the cal factors stored in the sensor
EEPROM. The command format is:
Syntax:
[AE or BE] KB [
n
] EN
[AE or BE] prefix specifies Sensor A or Sensor B
KB is the function code for Enter Cal Factor
[
n
] specifies a cal factor, expressed as a percentage with a valid range of 1.0% to 150.0%
EN is a required terminating suffix
Example:
AE KB 96 EN
! ENTERS A 96% CAL FACTOR FOR SENSOR A
Example 2:
BE KB 102 EN
! ENTERS A 102% CAL FACTOR FOR SENSOR B
Содержание 8650B Series
Страница 3: ...Series 8650B Universal Power Meters Publication 31470 001 Rev C November 2 2017 iii ...
Страница 18: ......
Страница 32: ......
Страница 78: ......
Страница 236: ......
Страница 244: ......
Страница 258: ......
Страница 263: ...Menu Structure Publication 31470 001 Rev C November 2 2017 D 3 Figure D 3 Modulation Sensor B Setup Menu Structure ...
Страница 265: ...Menu Stucture Publication 31470 001 Rev C November 2 2017 Page D 5 Figure D 4 Meter Setup Menu Structure ...
Страница 267: ...Publication 31470 001 Rev C November 2 2017 Last page of the document ...