Remote Operation
Publication 31470-001, Rev C, November 2, 2017
3-119
3.19
Advanced Features
3.19.1
Burst Start Exclude
Commands which cause the beginning of a burst to be excluded from measurement are based on the BSTE
function code (this feature is available only in the BAP mode):
Syntax:
[AE or BE] BSTE [a] EN
[AE or BE] prefix specifies Sensor A or Sensor B
BSTE along with the numeric variable [a] sets the number of burst start exclude samples in BAP
measurement mode
[a] specifies the number of 0.027 ms samples to be excluded; it has an integer value in the
range of 0 to 1686 with 1686 samples corresponding to 45.519 ms. The command is ignored for
[a] values that are out of acceptable range
Example:
AE BSTE 1 EN
! EXCLUDES ONE SAMPLE (= 0.027 mS) FROM START OF BURST
! FOR BAP MEASUREMENTS ON SENSOR A
Example 2:
BE BSTE 3 EN
! EXCLUDES THREE SAMPLES (= 0.081 mS) FROM START OF BURST
! FOR BAP MEASUREMENTS ON SENSOR B
3.19.2
Burst End Exclude
Commands which cause the end of a burst to be excluded from measurement are based on the BSPE function
code (this feature is available only in BAP mode):
Syntax:
[AE or BE] BSPE [b] EN
[AE or BE] prefix specifies Sensor A or Sensor B
BSPE along with the numeric variable [b] sets the number of burst end exclude samples in BAP
measurement mode
[b] specifies the number of 0.027 ms samples to be excluded; it has an integer value in the
range of 0 to 1183 (1183 samples correspond to 31.936 ms) when burst dropout time is 0.000
ms. For non-zero values of burst dropout time, the maximum value of end exclude time is limited
to 3.396 ms minus the burst dropout time setting. The command is ignored for [b] values that are
out of acceptable range
Example:
AE BSPE 1 EN
! EXCLUDES ONE SAMPLE (= 0.027 mS)FROM END OF BURST
! FOR BAP MEASUREMENTS ON SENSOR A
Example 2:
BE BSPE 3 EN
! EXCLUDES THREE SAMPLES (= 0.081 mS) FROM END OF BURST
! FOR BAP MEASUREMENTS ON SENSOR B
Содержание 8650B Series
Страница 3: ...Series 8650B Universal Power Meters Publication 31470 001 Rev C November 2 2017 iii ...
Страница 18: ......
Страница 32: ......
Страница 78: ......
Страница 236: ......
Страница 244: ......
Страница 258: ......
Страница 263: ...Menu Structure Publication 31470 001 Rev C November 2 2017 D 3 Figure D 3 Modulation Sensor B Setup Menu Structure ...
Страница 265: ...Menu Stucture Publication 31470 001 Rev C November 2 2017 Page D 5 Figure D 4 Meter Setup Menu Structure ...
Страница 267: ...Publication 31470 001 Rev C November 2 2017 Last page of the document ...