Series 8650B Universal Power Meters
4-4
Publication 31470-001, Rev C, November 2, 2017
4.3.2
Calibrator Frequency Check
To measure the frequency of the calibrator:
1.
Connect the frequency counter to the calibrator output connector.
2.
Turn ON the calibrator according to the procedure given in Procedure Step 8 on the previous page.
3.
Measure 49 to 51 MHz.
4.
Set the calibrator frequency to 1.000 GHz (with Option 12 only).
5.
Measure 999.5 MHz to 1.0005 GHz (with Option 12 only).
6.
Turn the calibrator OFF.
4.4
Performance Verification Tests
Information in this section is useful for periodic evaluation of the 8650B and its power sensors. These tests can
also be used for incoming inspection testing when the instrument is first received, if required.
If the 8650B has not previously been used, review the precautions in Section 1.2.1 of the publication before the
instrument is turned on. Prior to starting the following procedures, the instrument should be allowed to warm
up for at least 24 hours to assure maximum stability during testing.
The instrument plus power sensor linearity test is valid when the sensor has been calibrated using the front
panel calibrator at a temperature between 0° C and +50° C (+32° F to +122° F), and if operating within ±5°C (±9°
F) of that calibration temperature.
It is recommended that the verification be done in the order described since some of the steps use the
configuration from a previous step.
4.4.1
Equipment Required
See Table 4-1 for a list of the equipment required for performance testing of the 8650B power meters.
The Performance Verification Test Data Sheet located on page 4-8 through 4-9 can be printed out for recording
test results each time performance verification testing is performed on the specific instrument described by this
publication.
Содержание 8650B Series
Страница 3: ...Series 8650B Universal Power Meters Publication 31470 001 Rev C November 2 2017 iii ...
Страница 18: ......
Страница 32: ......
Страница 78: ......
Страница 236: ......
Страница 244: ......
Страница 258: ......
Страница 263: ...Menu Structure Publication 31470 001 Rev C November 2 2017 D 3 Figure D 3 Modulation Sensor B Setup Menu Structure ...
Страница 265: ...Menu Stucture Publication 31470 001 Rev C November 2 2017 Page D 5 Figure D 4 Meter Setup Menu Structure ...
Страница 267: ...Publication 31470 001 Rev C November 2 2017 Last page of the document ...