Performance Verification
Publication 31470-001, Rev C, November 2, 2017
4-9
SERIES 8650B UNIVERSAL POWER METERS
Performance Verification Test Data Sheet (Page 2 of 2)
Linearity Data -
(-60 dBm to +16 dBm)
Step
Attenuator
Value
Power Set
Point
Power
Meter
Reading
(P)
8650B
(DUT)
Reading
(R)
Reference
Power Ratio
8650B
(DUT)
Reading
Ratio
Linearity Error (%)
1
Linearity
Specification
Accumulated
Linearity
Error
2
0 dB
See Note 3
±
1%
Same as Lin.
error above
10 dB
1.00 mW
±
0.025 mW
P1 =
R1 =
P1/P2 =
R1/R2 =
10.0 mW
±
0.25 mW
P2 =
R2 =
±
1%
20 dB
1.00 mW
±
0.025 mW
P1 =
R1 =
P1/P2 =
R1/R2 =
10.0 mW
±
0.25 mW
P2 =
R2 =
±
1%
30 dB
1.00 mW
±
0.025 mW
P1 =
R1 =
P1/P2 =
R1/R2 =
10.0 mW
±
0.25 mW
P2 =
R2 =
±
1%
40 dB
1.00 mW
±
0.025 mW
P1 =
R1 =
P1/P2 =
R1/R2 =
10.0 mW
±
0.25 mW
P2 =
R2 =
±
1%
50 dB
1.00 mW
±
0.025 mW
P1 =
R1 =
P1/P2 =
R1/R2 =
10.0 mW
±
0.25 mW
P2 =
R2 =
±
1%
60 dB
1.00 mW
±
0.025 mW
P1 =
R1 =
P1/P2 =
R1/R2 =
10.0 mW
±
0.25 mW
P2 =
R2 =
±
1.5%
70 dB
1.00 mW
±
0.025 mW
P1 =
R1 =
P1/P2 =
R1/R2 =
10.0 mW
±
0.25 mW
P2 =
R2 =
±
6%
NOTES:
1.
Linearity Error (%) = [(R1/R2) / (P1/P2) - 1] x 100
2.
Accumulated error is the sum of the current dB segment linearity error plus the previous accumulated error.
3.
Use the first CW Linearity error value entered in the +16 dBm to +20 dBm Linearity Data on page 4-9.
Содержание 8650B Series
Страница 3: ...Series 8650B Universal Power Meters Publication 31470 001 Rev C November 2 2017 iii ...
Страница 18: ......
Страница 32: ......
Страница 78: ......
Страница 236: ......
Страница 244: ......
Страница 258: ......
Страница 263: ...Menu Structure Publication 31470 001 Rev C November 2 2017 D 3 Figure D 3 Modulation Sensor B Setup Menu Structure ...
Страница 265: ...Menu Stucture Publication 31470 001 Rev C November 2 2017 Page D 5 Figure D 4 Meter Setup Menu Structure ...
Страница 267: ...Publication 31470 001 Rev C November 2 2017 Last page of the document ...