Series 8650B Universal Power Meters
2-20
Publication 31470-001, Rev C, November 2, 2017
2.3.4
Sensor Setup
Sensor configuration enables selecting the operating mode of sensors, or to select the parameters for averaging
time-based measurements. From the Main Menu, press
[Sensor Setup]
to display the installed sensor data and
configuration options (See Figure 2-13). The power meter reads the contents of the EEPROMs for installed
sensors and displays each type of sensor and its connecting input.
The
[Sensor Setup]
softkey will not function if there are no sensors attached, or if attached sensors have not
been calibrated. The Sensor Setup options are dependent on the type of sensor. All sensors options are
Avg
and
Offset
. Peak and Modulation sensors add a
Config
option (See Figure 2-13).
Figure 2-13: Sensor Setup Menu
2.3.4.1
Averaging
Use this option to set the averaging period of power measurements, including modulated, CW and peak sensors.
Press
[Avg]
from the Sensor Setup menu to display the Averaging menu. This menu offers three options:
Auto
,
Averages
and
Time
.
Auto
1.
Select this setup option if the time is to be ‘adaptive’ averaged. That is, the time window changes
dynamically to reduce the effects of transient changes in power levels.
2.
Press
[OK]
to store the changes, or
[Cancel]
to exit without saving changes.
Averages
1.
Select this setup option to set the number of periods for averaging measurement signals over a fixed period
of time. Each unit is approximately 20 ms in duration. There is the choice of ‘powers of two’ multiples of
units from 1 to 1024. Set the number of periods with the up/down cursor keys.
2.
Press
[OK]
to store the changes, or
[Cancel]
to exit without saving changes.
Time
1.
Select this option to average measurement signals over a fixed period. A menu will display for to set the
time in seconds between 20 ms and 10 seconds. The meter will automatically change the selection to the
nearest of 20 ms times powers of 2 multiples of 1 to 1024 (e.g., entering 1.000 seconds will be changed to
1.280 seconds). Set the time digits with the cursor keys.
2.
Press
[OK]
to store the changes, or
[Cancel]
to exit without saving changes.
Содержание 8650B Series
Страница 3: ...Series 8650B Universal Power Meters Publication 31470 001 Rev C November 2 2017 iii ...
Страница 18: ......
Страница 32: ......
Страница 78: ......
Страница 236: ......
Страница 244: ......
Страница 258: ......
Страница 263: ...Menu Structure Publication 31470 001 Rev C November 2 2017 D 3 Figure D 3 Modulation Sensor B Setup Menu Structure ...
Страница 265: ...Menu Stucture Publication 31470 001 Rev C November 2 2017 Page D 5 Figure D 4 Meter Setup Menu Structure ...
Страница 267: ...Publication 31470 001 Rev C November 2 2017 Last page of the document ...