Remote Operation
Publication 31470-001, Rev C, November 2, 2017
3-93
3.11
Time Gating Measurement
The structure of the Time Gating remote commands for the Series 8650B Universal Power Meters has been
modified from the format used in the Series 8540C Universal Power Meters. The timing parameters cannot be
concatenated into a single command string. Each parameter must be specified in an individual string.
NOTE:
For the Series 8650B in 8540C Emulation, the 8540C Time Gating commands must be modified to
meet the 8650B Time Gating remote command structure.
3.11.1
Description
The time gating measurement option limits a power measurement to a defined interval that is controlled by a
start time and a duration. The start time begins after a programmable delay following a hardware trigger applied
to the Trigger Input connector on the instrument’s rear panel.
Off/Gated/Trigger/Edge
This selects between two modes of specifying the measurement period.
Off
Cancels the time gating function.
Gated
Selects the External Gating Mode in which measurements are taken while the trigger
input signal is true.
Trigger
Selects the External Trigger Mode and initiates a sequence of timers for defining the
measurement duration.
Edge
Selects the Internal Burst Detection Trigger Mode and initiates a sequence of timers
for defining the measurement duration.
3.11.1.1
External Gating Mode
Gate Polarity
This specifies the external input signal TTL high or low level as true for defining the
gated time.
3.11.1.2
External Trigger Mode
Trigger Polarity
This specifies the rising or falling edge of the external TTL trigger input signal as the
time reference point.
Trigger Delay
This is the delay time from receipt of the external trigger edge input to the start of the
gated measurement period.
Gate Time
This specifies of the length of the gated measurement period.
Holdoff Time
This is the timeout period between the end of the measurement period and the time
another trigger will be accepted.
3.11.1.3
Internal Burst Edge Detection
Trigger Delay
This is the delay time from the detection of the burst edge to the start of the gated
measurement period.
Gate Time
This specifies of the length of the gated measurement period.
Содержание 8650B Series
Страница 3: ...Series 8650B Universal Power Meters Publication 31470 001 Rev C November 2 2017 iii ...
Страница 18: ......
Страница 32: ......
Страница 78: ......
Страница 236: ......
Страница 244: ......
Страница 258: ......
Страница 263: ...Menu Structure Publication 31470 001 Rev C November 2 2017 D 3 Figure D 3 Modulation Sensor B Setup Menu Structure ...
Страница 265: ...Menu Stucture Publication 31470 001 Rev C November 2 2017 Page D 5 Figure D 4 Meter Setup Menu Structure ...
Страница 267: ...Publication 31470 001 Rev C November 2 2017 Last page of the document ...