Remote Operation
Publication 31470-001, Rev C, November 2, 2017
3-109
3.17.1.5
Warning Regarding Interruption & Reconfiguration
Another important consideration is that, while any of the three GPIB-fast measurement modes is running, it
should not be interrupted, and the measurement setup should not be changed. The measurement setup must
be completely configured before the command is sent to start the GPIB-fast measurement mode. To reconfigure
the instrument, or to zero a sensor, it is necessary to exit the GPIB-fast mode and then restart it. If a
measurement setup command is sent after a GPIB-fast mode command, the results are undefined.
3.17.1.6
GPIB-Fast Mode Setup
Prior to initiating a GPIB-fast measurement collection mode, the host should select the measurement (i.e., AP or
BP), select the measurement mode (i.e., CW, MAP, PAP, PEAK or BAP), define the frequency correction (via the
FR or KB command, but not via the V
PROP
F function), define the offset (if any), define the averaging (via the FA or
FM command), and define the duty cycle (if applicable). When a GPIB-fast mode is initiated, the display will
blank and a message will display indicating the GPIB-fast mode selected.
Содержание 8650B Series
Страница 3: ...Series 8650B Universal Power Meters Publication 31470 001 Rev C November 2 2017 iii ...
Страница 18: ......
Страница 32: ......
Страница 78: ......
Страница 236: ......
Страница 244: ......
Страница 258: ......
Страница 263: ...Menu Structure Publication 31470 001 Rev C November 2 2017 D 3 Figure D 3 Modulation Sensor B Setup Menu Structure ...
Страница 265: ...Menu Stucture Publication 31470 001 Rev C November 2 2017 Page D 5 Figure D 4 Meter Setup Menu Structure ...
Страница 267: ...Publication 31470 001 Rev C November 2 2017 Last page of the document ...