Remote Operation
Publication 31470-001, Rev C, November 2, 2017
3-117
3.18.3
PAP Mode
Commands which specify the Pulse Average Power measurement mode are based on the PAP function code:
Syntax:
PAP [A or B]
[A or B] specifies Sensor A or Sensor B
These commands will work only with a modulated sensor. Possible remote entry errors: 60/61 (uncalibrated or
missing sensor A/B), 62/63 (not a modulated sensor, or two sensor operation active, A/B). Refer to Table 3-22
for error numbers.
Example:
PAP A
! SELECTS PAP MODE FOR SENSOR A
Example 2:
PAP B
! SELECTS PAP MODE FOR SENSOR B
3.18.4
BAP Mode
Commands which specify the Burst Average Power measurement mode are based on the BAP function code:
Syntax:
BAP [A or B]
[A or BE] specifies Sensor A or Sensor B
These commands will work only with a modulated sensor. Possible remote entry errors: 60/61 (uncalibrated or
missing sensor A/B), 62/63 (not a modulated sensor, or two sensor operation active, A/B). Refer to Table 3-22
for error numbers.
Example:
BAP A
! SELECTS BAP MODE FOR SENSOR A
Example 2:
BAP B
! SELECTS BAP MODE FOR SENSORB
3.18.5
Peak Mode
The commands for Peak mode are discussed under separate headings for the 80350A sensors (See Section
Содержание 8650B Series
Страница 3: ...Series 8650B Universal Power Meters Publication 31470 001 Rev C November 2 2017 iii ...
Страница 18: ......
Страница 32: ......
Страница 78: ......
Страница 236: ......
Страница 244: ......
Страница 258: ......
Страница 263: ...Menu Structure Publication 31470 001 Rev C November 2 2017 D 3 Figure D 3 Modulation Sensor B Setup Menu Structure ...
Страница 265: ...Menu Stucture Publication 31470 001 Rev C November 2 2017 Page D 5 Figure D 4 Meter Setup Menu Structure ...
Страница 267: ...Publication 31470 001 Rev C November 2 2017 Last page of the document ...