Series 8650B Universal Power Meters
3-96
Publication 31470-001, Rev C, November 2, 2017
3.11.2.4
DELAY d
DELAY
d
specifies the delay time from the trigger to the start of the gating period. This parameter is relevant
only in External Trigger Mode and Burst Edge Detection Mode. Since the previous set up may be unknown, the
DELAY and DURATION (and HOLDOFF for External Trigger Mode) should be sent in series. Be aware that this
command is set up such that it automatically enables External Trigger Mode. That is, if this command is sent,
the External Trigger Mode will be enabled on the specified channel along with the specified trigger delay. To set
this parameter for Burst Edge Detection Mode, first send the specified trigger delay then enable the Burst Edge
Detection Mode with the GATE # EDGE command. For example, if the command is sent, the External Trigger
Mode will be enabled on channel A with a trigger delay of 20 ms.
Syntax:
GATE [A or B] DELAY [d]
GATE is the function code for the Timing Gate function.
[A or B] specifies Sensor A or Sensor B.
DELAY [d] in the External Trigger Mode, specifies the delay time from the trigger input edge to
the start of the gating period. In the Burst Edge Detection Mode, [d] specifies the delay time
from the detected burst edge to the start of the gating period. Variable [d] must be time specified
in seconds in the
range of 0 to 100 ms in 1 μs steps, where 0 represents some minimum non-
zero delay time.
Example:
GATE A DELAY 60E-6
! SPECIFIES A 60 µS DELAY FOR THE SENSOR A TIME GATE
! ALSO ENABLES EXTERNAL TRIGGER MODE
Example 2:
GATE B DELAY 20E-3
! SPECIFIES A 20 mS DELAY FOR THE SENSOR B TIME GATE
! ALSO ENABLES EXTERNAL TRIGGER MODE
3.11.2.5
DURATION g
DURATION
g
specifies the duration of the gating period. This parameter is relevant only in External Trigger Mode
and Burst Edge Detection Mode. Since the previous set up may be unknown, the DELAY and DURATION (and
HOLDOFF for External Trigger Mode) should be sent in series. Be aware that this command is set up such that it
automatically enables External Trigger Mode. That is, if this command is sent, the External Trigger Mode will be
enabled on the specified channel along with the specified gate duration. To set this parameter for Burst Edge
Detection Mode, first send the specified gate duration then enable the Burst Edge Detection Mode with the
GATE # EDGE command.
Syntax:
GATE [A or B] DURATION [g]
GATE is the function code for the Timing Gate function.
[A or B] specifies Sensor A or Sensor B.
DURATION [g] in the External Trigger Mode, specifies the duration of the gating period. In the
Burst Edge Detection Mode, [g] specifies the duration of the gating period. Variable [g] must be
a time specified in seconds in the range of 5 μs to 100 ms, with a resolution of 1 μs.
Example:
GATE A DURATION 680E-6
! SPECIFIES 680 µS GATE DURATION FOR THE SENSOR A TIME GATE
! ALSO ENABLES EXTERNAL TRIGGER MODE
Example 2:
GATE B DURATION 50E-3
! SPECIFIES 50 mS GATE DURATION FOR THE SENSOR B TIME GATE
! ALSO ENABLES EXTERNAL TRIGGER MODE
Содержание 8650B Series
Страница 3: ...Series 8650B Universal Power Meters Publication 31470 001 Rev C November 2 2017 iii ...
Страница 18: ......
Страница 32: ......
Страница 78: ......
Страница 236: ......
Страница 244: ......
Страница 258: ......
Страница 263: ...Menu Structure Publication 31470 001 Rev C November 2 2017 D 3 Figure D 3 Modulation Sensor B Setup Menu Structure ...
Страница 265: ...Menu Stucture Publication 31470 001 Rev C November 2 2017 Page D 5 Figure D 4 Meter Setup Menu Structure ...
Страница 267: ...Publication 31470 001 Rev C November 2 2017 Last page of the document ...