Series 8650B Universal Power Meters
3-12
Publication 31470-001, Rev C, November 2, 2017
3.2.8
Trigger Subsystem Commands
The TRIGger Subsystem is divided into two sections; Instrument Measurement Event Triggering and Special
Triggering Configuration commands for the GPIB fast-reading buffered data modes (GPIB Burst and GPIB Swift
modes). The TRIGger command tree is illustrated in Figure 3-5.
Figure 3-5: TRIGger subsystem Command Tree
The query form of these commands, TRIG?, will return the instrument’s current configuration status. This can be
used to verify triggering configuration or return status information following command errors which are
commonly caused by using illegal configuration commands during GPIB Swift or GPIB Burst modes.
SOURce:IMMediate triggering allows the 8650B to control measurement triggering; this is the default
configuration. EXTernal triggering is performed using a TTL signal input. BUS allows software controlled
triggering.
COUNt refers to the number of data points to store in the meter’s 5000 reading buffer, during GPIB Swift or
GPIB Burst mode, before the measurement data is requested by the controller.
DELay controls the time interval between GPIB Burst mode data samples, and the MODE command controls
whether the data is taken after receipt of the instrument trigger or if data is collected (in a FIFO buffer)
immediately preceding receipt of an instrument trigger.
Содержание 8650B Series
Страница 3: ...Series 8650B Universal Power Meters Publication 31470 001 Rev C November 2 2017 iii ...
Страница 18: ......
Страница 32: ......
Страница 78: ......
Страница 236: ......
Страница 244: ......
Страница 258: ......
Страница 263: ...Menu Structure Publication 31470 001 Rev C November 2 2017 D 3 Figure D 3 Modulation Sensor B Setup Menu Structure ...
Страница 265: ...Menu Stucture Publication 31470 001 Rev C November 2 2017 Page D 5 Figure D 4 Meter Setup Menu Structure ...
Страница 267: ...Publication 31470 001 Rev C November 2 2017 Last page of the document ...