Series 8650B Universal Power Meters
3-82
Publication 31470-001, Rev C, November 2, 2017
3.5.1.3
Freezing the Present Averaging Number
The command which causes auto filtering to hold its present averaging number is based on the FH function. The
command format is:
Syntax:
[AE or BE] FH
[AE or BE] prefix specifies sensor A or Sensor B
FH causes the 8650B to hold its present averaging number; auto averaging is deactivated
Example:
BE FH
! HOLDS PRESENT AVERAGE NUMBER FOR SENSOR B