Series 8650B Universal Power Meters
3-2
Publication 31470-001, Rev C, November 2, 2017
3.1.1.1
Clear Device over GPIB
The interface command
CLEAR 713
resets the GPIB and sets the 8650B to its preset condition.
3.1.1.2
Clear Interface over GPIB
The interface command
ABORT 7
resets the GPIB without resetting the 8650B to its preset condition. The 8650B
will not be addressed after the abort.
3.1.1.3
Local & Remote Control over GPIB
The interface command
LOCAL 713
places the 8650B into the local control mode.
The interface command
REMOTE 713
places the 8650B into the remote control mode. Enter LOCAL 713 to
return the instrument to local mode.
The interface command
LOCAL LOCKOUT 7
places the 8650B in the local lockout mode. This is a remote control
mode in which all of the 8650B front panel keys are disabled. The
GPIB LOCAL
command must be issued to
return the 8650B to local mode (Disconnecting the GPIB cable will also return the instrument to local mode).
3.1.1.4
Sensor Selection & Calibration
Power sensor selection data, specifications and calibration (local and remote) are contained in Appendix B of
this publication.
Содержание 8650B Series
Страница 3: ...Series 8650B Universal Power Meters Publication 31470 001 Rev C November 2 2017 iii ...
Страница 18: ......
Страница 32: ......
Страница 78: ......
Страница 236: ......
Страница 244: ......
Страница 258: ......
Страница 263: ...Menu Structure Publication 31470 001 Rev C November 2 2017 D 3 Figure D 3 Modulation Sensor B Setup Menu Structure ...
Страница 265: ...Menu Stucture Publication 31470 001 Rev C November 2 2017 Page D 5 Figure D 4 Meter Setup Menu Structure ...
Страница 267: ...Publication 31470 001 Rev C November 2 2017 Last page of the document ...