Remote Operation
Publication 31470-001, Rev C, November 2, 2017
3-23
3.2.12
Sensor Gate Commands
SENSe:GATE:MODE
SENSe:GATE:POLarity
SENSe:GATE:DELay
SENSe:GATE:DURation
SENSe:GATE:HOLDoff
The following commands are for sensor gate settings.
NOTE:
The Time Gate feature operates only with modulation sensors.
SENSe:GATE:MODE
Syntax:
SENSe<sensor 1 or 2>:GATE:MODE
space
<OFF, GATE, TRIGger or EDGE>
Example:
SENS1:GATE:MODE GATE
! SETS SENSOR 1 TO GATE TIME GATING MODE
Example 2:
SENS1:GATE:MODE?
! REQUESTS THE CURRENT SENSOR 1 GATE MODE
Description:
Configures a modulation sensor for the Time Gating operation. Selection is GATE, TRIGGER
or EDGE.
In GATE mode, an external TTL pulse supplied at the trigger input defines the measurement
gate. The meter measures the average power of a signal while the system is triggered. The
duration of the measurement is determined by the duration of the trigger signal.
In TRIGGER mode, an external TTL signal supplied at the trigger input initiates the trigger
cycle. After the set Gate Delay time has been satisfied, the meter proceeds to measure the
average power of a signal for the duration specified in by the Gate Duration setting.
In EDGE mode, the meter internally detects the rising edge of a pulsed signal to trigger a gated
measurement. After the set Gate Delay time has been satisfied, the meter proceeds to
measure the average power of a signal for the duration specified in by the Gate Duration
setting.
Query form for sensor Time Gate mode is available. Response can be OFF, GATE, TRIGGER
or EDGE.
SENSe:GATE:POLarity
Syntax:
SENSe<sensor 1 or 2>:GATE:POLarity
space
<INVert or NINVert>
Example:
SENS1:GATE:POL NINV
! SETS NON-INVERTED POLARITY FOR THE SENSOR 1 GATE
TRIGGER INPUT
Example 2:
SENS1:GATE:POL?
! REQUESTS SENSOR 1 GATE TRIGGER POLARITY SETTING
Description:
Configures the Time Gating trigger input for positive or negative polarity. NINVert configures
the meter to trigger from the rising edge of the trigger input while INVert configures the meter to
trigger from the falling edge of the trigger input.
Query form for sensor Time Gate polarity is available. Response can be INVERT or NINVERT.
Содержание 8650B Series
Страница 3: ...Series 8650B Universal Power Meters Publication 31470 001 Rev C November 2 2017 iii ...
Страница 18: ......
Страница 32: ......
Страница 78: ......
Страница 236: ......
Страница 244: ......
Страница 258: ......
Страница 263: ...Menu Structure Publication 31470 001 Rev C November 2 2017 D 3 Figure D 3 Modulation Sensor B Setup Menu Structure ...
Страница 265: ...Menu Stucture Publication 31470 001 Rev C November 2 2017 Page D 5 Figure D 4 Meter Setup Menu Structure ...
Страница 267: ...Publication 31470 001 Rev C November 2 2017 Last page of the document ...