Series 8650B Universal Power Meters
3-128
Publication 31470-001, Rev C, November 2, 2017
3.19.9
Preset
The PR command resets the 8650B to its default settings, leaving the user settings intact as
Previous Settings
if
they were different from the default settings. This command does not function while in the SWIFT, FBUF or
FMOD modes.
Alternatively, the IEEE 488.2 command *RST also resets the 8650B to its default settings, and functions in the
SWIFT, FBUF and FMOD modes. These modes must be re-entered over the GPIB.
The preset conditions of the instrument are outlined in Table 3-18.
Table 3-18: Preset (Default) Conditions
Sensors
(All parameters apply to Sensor A & Sensor B)
General
Parameter
Condition
Parameter
Condition
Cal Factor
100.0%
Sensor Selection
Sensor A
Offset
0.00 dB
Calibrator
Off
Filter
AUTO
Default Sensor Prefix
Sensor A
Range
AUTO
Resolution
2 (0.01 dB)
Low Limit
0.000 dBm
Limits Checking
Off
High Limit
0.000 dBm
Max/Min
Off
Frequency
50 MHz
REL
Off
Duty Cycle
OFF, 1.000%
Trigger Mode
TR3
Measurement Mode
See Note 1
Group Execute Trigger Mode
GT2
Display Function
Display Enable
Peaking Meter Mode
Status
Pulse Sensor Mode
Internal Trigger
Measurement Units
See Note 2
Sound
On
Analog Output
Off
Notes:
1.
The default measurement mode depends on the sensor type. For a CW sensor, the default is CW. For a
modulation sensor, the default is MAP. For a peak sensor, the default is INT TRIG. Regardless of the
sensor type, all advanced features are turned off.
2.
There is a slight difference between the preset conditions as set by a remote command, and as set from
the front panel menus. This difference has to do with measurement units. If the 8650B is preset from
the front panel, this sets the measurement units to dBm in all cases. If the 8650B is preset over the bus,
this sets the measurement units to Watts in the case of HP438 emulation, and has no effect at all in the
case of HP436 emulation; otherwise, it sets the measurement units to dBm.
This distinction is made to accommodate differences between emulations for remote programming
purposes without affecting the benchtop user.
Содержание 8650B Series
Страница 3: ...Series 8650B Universal Power Meters Publication 31470 001 Rev C November 2 2017 iii ...
Страница 18: ......
Страница 32: ......
Страница 78: ......
Страница 236: ......
Страница 244: ......
Страница 258: ......
Страница 263: ...Menu Structure Publication 31470 001 Rev C November 2 2017 D 3 Figure D 3 Modulation Sensor B Setup Menu Structure ...
Страница 265: ...Menu Stucture Publication 31470 001 Rev C November 2 2017 Page D 5 Figure D 4 Meter Setup Menu Structure ...
Страница 267: ...Publication 31470 001 Rev C November 2 2017 Last page of the document ...