Series 8650B Universal Power Meters
3-102
Publication 31470-001, Rev C, November 2, 2017
3.14.1
Learn Mode #1
Learn Mode #1 is used to return the configuration of the 8650B to the controller in the form of a sequence of
remote commands.
3.14.2
Requesting the String
The simple command which requests the Learn Mode #1 string has the following format:
Syntax:
LP1
Example:
LP1
! REQUESTS LEARN MODE #1 STRING
Response:
TR3APAEKB100.0ENOS+000.00ENRA 000.000.000ENBEKB100.0ENOS+000.00ENRA ENFA
LL+000.000.000ENAELGOC0GT0AELM0BELM0<CR><LF>
After receiving the LP1 command, the 8650B will return the Learn Mode #1 string the next time it is addressed
to talk. The string will consist of up to 128 ASCII characters. The last character is sent with EOI true. Table 3-17
shows the information contained in the Learn Mode #1 string and the order in which it is sent.
Table 3-17: Learn Mode #1 Output Format
Parameter
Output from the Power Meter
1
Trigger Mode
TRd
Measurement Mode
AP, BP, AR, BR, AD, or BD
SENSOR A PARAMETERS
Cal Factor
Offset
Range
Filter
Low Limit
High Limit
AE
KB ddd.d EN
OS ±dd.dd EN
RA d EN
FA or FM d EN
LL ±ddd.ddd EN
LH ±ddd.ddd EN
SENSOR B PARAMETERS
Cal Factor
Offset
Range
Filter
Low Limit
High Limit
BE
KB ddd.d EN
OS ±dd.dd EN
RA d EN
FA or FM d EN
LL ±ddd.ddd EN
LH ±ddd.ddd EN
Active Entry Channel
AE or BE
Measurement Units
LG or LN
Reference Oscillator Status
OC0 or OC1
Group Trigger Mode
GTd
Sensor A Limits Checking Status
AE LM0 or AE LM1
Sensor B Limits Checking Status
BE LM0 or BE LM1
Carriage Return <CR>, Line Feed <LF>
EOI
1: ± indicates sign; d indicates a single digit.
3.14.2.1
Sending the String
The power meter can be restored to the configuration described in the Learn Mode #1 string, by sending the
string to the 8650B.
Содержание 8650B Series
Страница 3: ...Series 8650B Universal Power Meters Publication 31470 001 Rev C November 2 2017 iii ...
Страница 18: ......
Страница 32: ......
Страница 78: ......
Страница 236: ......
Страница 244: ......
Страница 258: ......
Страница 263: ...Menu Structure Publication 31470 001 Rev C November 2 2017 D 3 Figure D 3 Modulation Sensor B Setup Menu Structure ...
Страница 265: ...Menu Stucture Publication 31470 001 Rev C November 2 2017 Page D 5 Figure D 4 Meter Setup Menu Structure ...
Страница 267: ...Publication 31470 001 Rev C November 2 2017 Last page of the document ...