Remote Operation
Publication 31470-001, Rev C, November 2, 2017
3-53
3.2.25
Min/Max Configuration & Monitoring
CALCulate:MAXimum
CALCulate:MINimum
MINimum and MAXimum channel monitoring records the minimum and maximum variation of channel
amplitude over time. Time zero is set by setting the MINimum or MAXimum STATe to ON. The default preset
and turn on configuration is OFF. After setting to ON, the minimum and maximum value of the measured value
on that channel will be recorded. To reset the MINimum and MAXimum value back to the current value, set the
STATe to ON again. The STATe OFF does not need to be sent to reset of MINimum and MAXimum values.
Individual sensor minimum and maximum power level variations can be monitored only if that sensor is defined
directly to a channel with POW 1 or POW 2. MINimum and MAXimum channel monitoring is only available in
NORMal mode. GPIB BURSt and GPIB SWIFt modes do not perform MINimum and MAXimum monitoring.
CALCulate:MAXimum
Syntax:
CALCulate<channel 1 to 4>:MAXimum:STATe
space
<ON | OFF>
Example:
CALC2:MAX:STAT ON
! ENABLES CHANNEL 2 MAXIMUM TRACKING
Description:
This command enables and disables the MAXimum monitoring function on the specified
channel. The maximum value of the channel (not sensor) is monitored until the value is reset by
CALC#:MAX:STAT ON or CALC#:MAX:STAT OFF is sent.
Syntax:
CALCulate<channel 1 to 4>:MAXimum[:MAGnitude]?
Example:
CALC2:MAX?
! QUERIES CHANNEL 2 MAXIMUM VALUE IN DBM
Response:
Highest power reading since CALC2:MAX:STAT ON was sent or reset.
Description:
This command reports the value maintained as the maximum in the 8650B max monitor.
CALCulate:MINimum
Syntax:
CALCulate<channel 1 to 4>:MINimum:STATe
space
<ON | OFF>
Example:
CALC1:MIN:STAT ON
! ENABLES CHANNEL 1 MINIMUM TRACKING
Description:
This command enables and disables the MINimum monitoring function on the specified
channel. The minimum value of the channel (not sensor) is monitored until the value is reset by
CALC#:MIN:STAT ON or CALC#:MIN:STAT OFF is sent.
Syntax:
CALCulate<channel 1 to 4>:MINimum[:MAGnitude]?
Example:
CALC1:MIN?
! QUERIES CHANNEL 1 MINIMUM VALUE IN DBM
Response:
Lowest power reading since CALC1:MIN:STAT ON was sent or reset.
Description:
This command reports the value maintained as the minimum in the 8650B min monitor.
Содержание 8650B Series
Страница 3: ...Series 8650B Universal Power Meters Publication 31470 001 Rev C November 2 2017 iii ...
Страница 18: ......
Страница 32: ......
Страница 78: ......
Страница 236: ......
Страница 244: ......
Страница 258: ......
Страница 263: ...Menu Structure Publication 31470 001 Rev C November 2 2017 D 3 Figure D 3 Modulation Sensor B Setup Menu Structure ...
Страница 265: ...Menu Stucture Publication 31470 001 Rev C November 2 2017 Page D 5 Figure D 4 Meter Setup Menu Structure ...
Страница 267: ...Publication 31470 001 Rev C November 2 2017 Last page of the document ...