Remote Operation
Publication 31470-001, Rev C, November 2, 2017
3-27
3.2.14
Instrument Triggering
*TRG
TRIGger[:IMMediate]
TRIGger:SOURce
TRIGger:MODE
TRIGger:DELay
TRIGger:COUNt
*WAI
These SCPI commands trigger the measurement cycle. They do not configure or provide triggering for Peak
Power Sensors. Those commands are defined in Section 3.2.19. For power meter operation, the TRIGger
Subsystem is divided into two sections; Instrument Measurement Event Triggering, and Special Triggering
Configuration commands for the fast-reading buffered data: GPIB Burst and GPIB Swift modes.
Figure 3-8: TRIGer Sub-System Command Tree
TRIGger:SOURce IMMediate triggering allows the 8650B to free-run and perform continuous measurements.
This is the default 8650B trigger setting. During NORMal mode, with TRIGger:SOURce set to IMMediate, power
measurements can be read with MEASure?, READ? (when preceeded with INITiate:IMMediate), or FETCh?
(when preceeded with INITitiate:CONTinuous ON). TRIGger:SOURce IMMediate triggering is not compatible with
the GPIB BURSt and GPIB SWIFt modes. If sending the CALC#:MODE BURS command to enter GPIB BURSt mode,
the instrument TRIGger:SOURce will automatically be switched from IMMediate to BUS. If sending the
CALC#:MODE SWIF command to enter GPIB SWIFt mode, the instrument TRIGger:SOURce source will remain
IMMediate but a device-specific error will be generated whenever specifying a TRIG:COUN # higher than 1. As a
general rule, when a measurement subroutine switches operating modes, proceed to send the required
triggering configuration as well.
TRIGger:SOURce EXTernal triggering is performed with the rear panel Trigger Input BNC connector and functions
only in the GPIB BURSt and GPIB SWIFt modes. EXTernal triggering is not available in the NORMal mode.
TRIGger:SOURce BUS triggering sets the triggering control to the controller software using TRIG or *TRG
commands and is available for all operating modes: GPIB BURSt, GPIB SWIFt and NORMal.
Содержание 8650B Series
Страница 3: ...Series 8650B Universal Power Meters Publication 31470 001 Rev C November 2 2017 iii ...
Страница 18: ......
Страница 32: ......
Страница 78: ......
Страница 236: ......
Страница 244: ......
Страница 258: ......
Страница 263: ...Menu Structure Publication 31470 001 Rev C November 2 2017 D 3 Figure D 3 Modulation Sensor B Setup Menu Structure ...
Страница 265: ...Menu Stucture Publication 31470 001 Rev C November 2 2017 Page D 5 Figure D 4 Meter Setup Menu Structure ...
Страница 267: ...Publication 31470 001 Rev C November 2 2017 Last page of the document ...