Front Panel Operation
Publication 31470-001, Rev C, November 2, 2017
2-9
Sound
A speaker within the chassis produces audible clicks and tones to register keystrokes and to draw attention to
certain conditions. For example, if a limit has been exceeded or a calibration process has been completed. From
the Power Meter Configuration menu, move the cursor to the
Sound
option and press
[Config]
. The Sound
Config menu will display. Press
[On]
to enable the speaker, or
[Off]
to disable it. There is no sub menu
associated with this feature.
RS-232
From the Power Meter Configuration menu, move the cursor to the
RS-232
option and press
[Config]
. The RS-
232 Config menu will display for configuration of the RS-232 interface.
1.
Press
[Baud]
to set the Baud Rate. Select the desired rate with the up/down cursor keys. The default
value is
9600
.
2.
Press
[Length]
to set the Data Bits. Select the desired data bit length with the up/down cursor keys. The
default value is
8
.
3.
Press
[Parity]
and set the parity with the up/down cursor keys. The default parity is set to
None
.
4.
Press
[Stop bits]
to set the stop bit value. Select the desired stop bit value with the up/down cursor
keys. The default value is
1
.
5.
Press
[OK]
to accept the changes or
[Cancel]
to cancel the changes and return to the Power Meter
Configuration menu.
Содержание 8650B Series
Страница 3: ...Series 8650B Universal Power Meters Publication 31470 001 Rev C November 2 2017 iii ...
Страница 18: ......
Страница 32: ......
Страница 78: ......
Страница 236: ......
Страница 244: ......
Страница 258: ......
Страница 263: ...Menu Structure Publication 31470 001 Rev C November 2 2017 D 3 Figure D 3 Modulation Sensor B Setup Menu Structure ...
Страница 265: ...Menu Stucture Publication 31470 001 Rev C November 2 2017 Page D 5 Figure D 4 Meter Setup Menu Structure ...
Страница 267: ...Publication 31470 001 Rev C November 2 2017 Last page of the document ...