Remote Operation
Publication 31470-001, Rev C, November 2, 2017
3-21
3.2.11
Sensor Configuration Commands
SENSe:CONFig?
SENSe:CONFig:CW
SENSe:CONFig:MAP
SENSe:CONFig:MAP:UNSYnchronized
SENSe:CONFig:BAP
SENSe:CONFig:PAP
The following commands are for sensor configuration.
SENSe:CONFig?
Syntax:
SENSe<sensor 1 or 2>:CONFig?
Example:
SENS1:CONF?
! QUERIES SENSOR 1 CONFIGURATION
Description:
Queries sensor configuration on specified input. Response is character response data:
CW | MAP | MAPU | BAP | PAP | PEAK |UNCAL | NOSENSOR.
SENSe:CONFig:CW
Syntax:
SENSe<sensor 1 or 2>:CONFig:CW
Example:
SENS1:CONF:CW
! ENABLES CW MODE FOR SENSOR 1
Description:
Configures a modulation or peak sensor for CW mode on the specified input.
SENSe:CONFig:MAP
Syntax:
SENSe<sensor 1 or 2>:CONFig:MAP[:SYNChronized]
Example:
SENS1:CONF:MAP
! ENABLES SYNCHRONIZED MAP MODE FOR SENSOR 1
Description:
Configures a modulation sensor for Synchronized MAP (Modulated Average Power) mode on
the specified input.
SENSe:CONFig:MAP:UNSYnchronized
Syntax:
SENSe<sensor 1 or 2>:CONFig:MAP:UNSYnchronized
Example:
SENS1:CONF:MAP:UNSY
! ENABLES UNSYNCHRONIZED MAP MODE FOR SENSOR 1
Description:
Configures a modulation sensor for Unsynchronized MAP (Modulated Average Power) mode on
the specified input. This mode is effective when performing measurements on signals with high
average values where the signal exhibits very large power swings. If such an irregularly
modulated signal is measured using Synchronized MAP mode, measurement settling time will
vary as the power meter attempts to synchronize to the modulation. Disabling synchronization
with Unsynchronized MAP mode will achieve faster measurement results on such a signal.
Содержание 8650B Series
Страница 3: ...Series 8650B Universal Power Meters Publication 31470 001 Rev C November 2 2017 iii ...
Страница 18: ......
Страница 32: ......
Страница 78: ......
Страница 236: ......
Страница 244: ......
Страница 258: ......
Страница 263: ...Menu Structure Publication 31470 001 Rev C November 2 2017 D 3 Figure D 3 Modulation Sensor B Setup Menu Structure ...
Страница 265: ...Menu Stucture Publication 31470 001 Rev C November 2 2017 Page D 5 Figure D 4 Meter Setup Menu Structure ...
Страница 267: ...Publication 31470 001 Rev C November 2 2017 Last page of the document ...