Table 3-13: HP437 Emulation Command Set
Command
Description
LN
Linear display
LP2
HP437 learn mode
LT
Left arrow [ignored]
OC0
Reference oscillator off
OC1
Reference oscillator on
OD
Output display text [ignored]
OF0
Offset off – Local
OF1
Offset on – Local
OS
Offset (enter offset value)
PR
Preset
RA
Auto range
4
RC
Recall
1
RE
Resolution
1
RF0 - RF9
Enter sensor ref cal factor [ignored]
RH
Range hold
4
RL0
Exit REL mode
RL1
Enter REL mode using REL value
RL2
Use old ref number
RM
Set range
1, 4
RT
Right arrow [ignored]
RV
Read Service Request Mask value
SE
Sensor [ignored]
SM
Status Message
SN0 - SN9
enter sensor serial number [ignored]
ST
Store instrument state
TR0
Trigger hold
TR1
Trigger immediate
TR2
Trigger with delay
TR3
Trigger - free run
UP
Up arrow [up arrow]
ZE
Zero
Notes:
1.
A numeric entry is required by these GPIB codes, followed by the code EN (ENTER).
2.
This GPIB code uses the next 6 characters (0-9, A-Z, or an underscore) as input data.
3.
The asterisk (*) must be included as part of the GPIB command string.
4.
The 8650B can always measure over its entire dynamic range; there is no need to specify the range.
Therefore, range-related commands have no effect on the measurement capability of the 8650B. The
auto range, range hold, and set range commands only offset the analog output voltage, and only in
HP436, HP437, or HP438 GPIB emulation modes. In these emulation modes (when using a single sensor,
and not measuring in a relative mode), the power will be scaled to a range of 0 to 1 volts, representing
the relative power within the current 10 dB range of the 8650B. The range hold and set range
commands will simulate locking the range of power represented by the output voltage.
Содержание 8650B Series
Страница 3: ...Series 8650B Universal Power Meters Publication 31470 001 Rev C November 2 2017 iii ...
Страница 18: ......
Страница 32: ......
Страница 78: ......
Страница 236: ......
Страница 244: ......
Страница 258: ......
Страница 263: ...Menu Structure Publication 31470 001 Rev C November 2 2017 D 3 Figure D 3 Modulation Sensor B Setup Menu Structure ...
Страница 265: ...Menu Stucture Publication 31470 001 Rev C November 2 2017 Page D 5 Figure D 4 Meter Setup Menu Structure ...
Страница 267: ...Publication 31470 001 Rev C November 2 2017 Last page of the document ...