Remote Operation
Publication 31470-001, Rev C, November 2, 2017
3-1
Chapter 3
Remote Operation
3.1
Introduction
The Series 8650B can be operated from a remote host over the General Purpose Interface Bus (GPIB), Universal
Serial Bus (USB), or Local Area Network (LAN) using either Standard Commands for Programmable Instruments
(SCPI) or IEEE Standard 488-1978 (Digital Interface for Programmable Instruments) commands. Table 3-1 shows
which functions of the IEEE 488 standards are implemented in the 8650B.
Table 3-1: Implemented IEEE Standards
Function
8650B Implementation
Source Handshake
SH1 (complete capability)
Acceptor Handshake
AH1 (complete capability)
Talker
T5 (basic talker, serial poll, talk only mode, unaddressed if MLA)
Extended Talker
TE0 (no capability)
Listener
L3 (basic listener, listen only mode, unaddressed if MTA)
Extended Listener
LE0 (no capability)
Service Request
SR1 (complete capability)
Remote/Local
RL1 (complete capability)
Parallel Poll
PP1 (remote configuration)
Device Clear
DC1 (complete capability)
Device Trigger
DT1 (complete capability)
Controller
C0 (no capability)
3.1.1
Sending Commands to the 8650B over GPIB
The 8650B power meter uses standard protocols for communication over the GPIB. Commands conform to IEEE
488.1 or IEEE 488.2 guidelines. Four emulation modes (Giga-tronics 8540, HP436, HP437 and HP438) are
available for users of power meters who cannot rewrite their application software.
The program examples in this chapter are written in HTBasic™ format (HTBasic is a trademark of TransEra
Corporation). Other languages would use different commands but the string that is sent or received will always
be the same. In HTBasic, the
OUTPUT
command sends a string to the GPIB. The number after
OUTPUT
is the
GPIB address of the instrument.
The factory-set default address of the 8650B is 13 and the address of the GPIB is assumed to be 7; therefore,
examples of command strings in this publication are preceded by
OUTPUT 713
.
Содержание 8650B Series
Страница 3: ...Series 8650B Universal Power Meters Publication 31470 001 Rev C November 2 2017 iii ...
Страница 18: ......
Страница 32: ......
Страница 78: ......
Страница 236: ......
Страница 244: ......
Страница 258: ......
Страница 263: ...Menu Structure Publication 31470 001 Rev C November 2 2017 D 3 Figure D 3 Modulation Sensor B Setup Menu Structure ...
Страница 265: ...Menu Stucture Publication 31470 001 Rev C November 2 2017 Page D 5 Figure D 4 Meter Setup Menu Structure ...
Страница 267: ...Publication 31470 001 Rev C November 2 2017 Last page of the document ...