Remote Operation
Publication 31470-001, Rev C, November 2, 2017
3-19
3.2.10
Sensor Calibration & Zeroing
CALibrate
CALibrate:STATe?
CALibrate:ZERO
SENSe:TEMPerature?
3.2.10.1
Sensor Calibration
The CALibration commands for sensor calibration and zeroing are important for accurate power measurement
results. Be sure to perform the sensor calibration prior to beginning measurement operation or channel
configuration. Sensors must be calibrated to the meter before performing measurements.
Zeroing of all active sensors should always be performed whenever a second sensor (whether calibrated or not)
is added or removed. Zeroing should also be performed prior to measurement of low signal levels, generally
within the lower 15 dB of a sensor’s dynamic range. For standard sensors, this is -55 dBm.
CALibrate
Syntax:
CALibration<sensor 1 or 2>[:SENSor]
Example:
CAL1
! CALIBRATE SENSOR 1
Description:
This command begins the sensor power sweep calibration process. Power sweep calibration
can be performed only during Normal mode. The sensor must be attached to the front panel
CALIBRATOR (use an adapter for high frequency sensors with Type K or 2.4mm connectors).
If the sensor is not connected or if the sensor is disconnected during the power sweep
calibration procedure, the calibration automatically fails and the sensor power sweep
calibration table is restored to its previous values.
CALibrate:STATE?
Syntax:
CALibrate<sensor 1 or 2>:STATe?
Example:
CAL1:STAT?
! QUERIES SENSOR FOR CALIBRATION STATUS
Response:
1 if sensor is calibrated or 0 if sensor is not calibrated.
Description:
This command queries whether or not a sensor has been calibrated.
Содержание 8650B Series
Страница 3: ...Series 8650B Universal Power Meters Publication 31470 001 Rev C November 2 2017 iii ...
Страница 18: ......
Страница 32: ......
Страница 78: ......
Страница 236: ......
Страница 244: ......
Страница 258: ......
Страница 263: ...Menu Structure Publication 31470 001 Rev C November 2 2017 D 3 Figure D 3 Modulation Sensor B Setup Menu Structure ...
Страница 265: ...Menu Stucture Publication 31470 001 Rev C November 2 2017 Page D 5 Figure D 4 Meter Setup Menu Structure ...
Страница 267: ...Publication 31470 001 Rev C November 2 2017 Last page of the document ...