Series 8650B Universal Power Meters
1-10
Publication 31470-001, Rev C, November 2, 2017
Instrumentation Linearity (continued)
Zero Drift
2,4
<±100 pW with all CW Power Sensors
<±200 pW with Series 804XXA and 806XXA Modulation Power
Sensors
<±400 pW with Series 807XXA Modulation Power Sensors
Noise
2,5
<±50 pW with all CW Power Sensors
<±100 pW with Series 804XXA and 806XXA Modulation Power
Sensors
<±200 pW with Series 807XXA Modulation Power Sensors
Notes:
1.
Depending on sensor used (See Power Sensor details in Appendix B).
2.
Specified performance applies with maximum averaving and 24 hour warm-up at constant temperature.
3.
Measurable over any 1-minute interval after zeroing, three standard deviations.
4.
Measurable over any 1-hour interval after zeroing, three standard deviations.
Zero Drift Measurement:
a.
Set the meters Average to 1024. Perform Calibration. Connect a 50-ohm load to the sensor after
Calibration and Zeroing.
b.
Temperature stabilize at 25 °C for 24 hours.
c.
After the 24-hour stabilization, perform a Zero Drift measurement.
d.
Zero the meter and take an initial measurement reading in linear measurement units.
e.
Continue taking one reading every 10 minutes until 6 readings have been taken.
Plot the 6 readings; Zero Drift should be ±100pW, ±200pW or ±400pW, depending on the sensor.
5.
Measurable over any 1-minute interval at constant power, three standard deviations.
Содержание 8650B Series
Страница 3: ...Series 8650B Universal Power Meters Publication 31470 001 Rev C November 2 2017 iii ...
Страница 18: ......
Страница 32: ......
Страница 78: ......
Страница 236: ......
Страница 244: ......
Страница 258: ......
Страница 263: ...Menu Structure Publication 31470 001 Rev C November 2 2017 D 3 Figure D 3 Modulation Sensor B Setup Menu Structure ...
Страница 265: ...Menu Stucture Publication 31470 001 Rev C November 2 2017 Page D 5 Figure D 4 Meter Setup Menu Structure ...
Страница 267: ...Publication 31470 001 Rev C November 2 2017 Last page of the document ...