Series 8650B Universal Power Meters
3-142
Publication 31470-001, Rev C, November 2, 2017
3.19.16
Units
NOTE:
These commands are specific to the active Channel. These commands must be preceded by the
appropriate Channel Configuration using the CH [n] EN command. See Section 3.7.4 for details.
Logarithmic or linear measurement units are specified by simple commands consisting of the function codes LG
and LN.
Syntax:
LG or LN
Example:
LG
! SETS LOG UNITS (DB OR DBM) FOR THE ACTIVE CHANNEL
Example 2:
LN
! SETS LINEAR UNITS (WATTS OR %) FOR THE ACTIVE CHANNEL
These commands affect all types of measurements, except for the GPIB-fast measurement collection modes.
These modes always return measurement readings in dBm.
3.19.17
V
PROP
F Feature
The V
PROP
F feature (Voltage Proportional to Frequency) provides a means of indicating to the 8650B the
approximate frequency of the signal that it is measuring, so that the appropriate cal factor can be applied. The
frequency is indicated by means of a variable voltage input. The 8650B reads the voltage as an expression of
frequency, and applies the proper cal factor from the table stored in the sensor EEPROM (interpolating for
frequencies that fall between the stored values).
In order for the 8650B to interpret the input voltage input correctly, it is necessary to specify the starting point
(that is, the frequency at zero volts) and the slope (the rate at which voltage increases with frequency).
3.19.17.1
Enabling & Disabling V
PROP
F
Commands related to the V
PROP
F function are based on the function code VPROPF. The command format, for
purposes of activating or deactivating the V
PROP
F feature, is as follows:
Syntax:
VPROPF [A or B] STATE [ON | OFF]
VPROPF is the function code for the V
PROP
F function
[A or B] specifies Sensor A or Sensor B
STATE indicates that the V
PROP
F feature is being enabled or disabled
[ON | OFF] enables or disables the V
PROP
F function
Example:
VPROPF A STATE ON
! ENABLES V
PROP
F FOR SENSOR A
Example 2:
VPROPF B STATE OFF
! DISABLES V
PROP
F FOR SENSOR B
Содержание 8650B Series
Страница 3: ...Series 8650B Universal Power Meters Publication 31470 001 Rev C November 2 2017 iii ...
Страница 18: ......
Страница 32: ......
Страница 78: ......
Страница 236: ......
Страница 244: ......
Страница 258: ......
Страница 263: ...Menu Structure Publication 31470 001 Rev C November 2 2017 D 3 Figure D 3 Modulation Sensor B Setup Menu Structure ...
Страница 265: ...Menu Stucture Publication 31470 001 Rev C November 2 2017 Page D 5 Figure D 4 Meter Setup Menu Structure ...
Страница 267: ...Publication 31470 001 Rev C November 2 2017 Last page of the document ...