Series 8650B Universal Power Meters
3-86
Publication 31470-001, Rev C, November 2, 2017
3.7.2
Calibrator Source
The 8650B Calibrator output (a fixed signal at 0 dBm) is activated and deactivated by means of two simple
commands:
Syntax:
[OC1 or OC0]
[OC1 or OC0] turns ON and OFF the calibration source. This command resets the calibration
output to 50 MHz, 0.00 dBm.
Example:
OC1
! TURNS ON CALIBRATOR SOURCE
Example 2:
OC0
! TURNS OFF CALIBRATOR SOURCE
NOTE:
This command is needed for test purposes only. The calibrator source is enabled automatically
during calibration of a sensor.
The following command sets the calibrator frequency for Option 12:
Syntax:
[OC1]
d
[HZ or KZ or MZ or GZ]
OC1 turns ON the calibration source
d
is the frequency setting of the calibrator output whose value is determined by the specified
numerical variable together with the specified unit suffix. The range is 50 MHz and 800 MHz to
1 GHz in 0.5 MHz increments
[HZ or KZ or MZ or GZ] is the required terminating unit suffix
Example:
OC1 1.000 GZ
! SETS THE CALIBRATOR FREQUENCY TO 1.000 GHZ
Example 2:
OC1 950 MZ
! SETS THE CALIBRATOR FREQUENCY TO 950 MHZ
NOTE:
This command resets calibration output to 0.00 dBm.
Содержание 8650B Series
Страница 3: ...Series 8650B Universal Power Meters Publication 31470 001 Rev C November 2 2017 iii ...
Страница 18: ......
Страница 32: ......
Страница 78: ......
Страница 236: ......
Страница 244: ......
Страница 258: ......
Страница 263: ...Menu Structure Publication 31470 001 Rev C November 2 2017 D 3 Figure D 3 Modulation Sensor B Setup Menu Structure ...
Страница 265: ...Menu Stucture Publication 31470 001 Rev C November 2 2017 Page D 5 Figure D 4 Meter Setup Menu Structure ...
Страница 267: ...Publication 31470 001 Rev C November 2 2017 Last page of the document ...