Series 8650B Universal Power Meters
3-80
Publication 31470-001, Rev C, November 2, 2017
3.4
Analog Output
These commands control the A and B analog outputs.
3.4.1
Enabling & Disabling the Output
The ANALOG function can enable or disable the analog outputs. The command format for this purpose is:
Syntax:
ANALOG [STD or OPT] STATE [ON | OFF]
ANALOG is the function code for Configure Analog Output
[STD or OPT] specifies analog output A or B respectively
STATE indicates that the analog output ON/OFF status is being configured
[ON | OFF] specifies whether the analog output is to be enabled or disabled
Example:
ANALOG STD STATE ON
! ENABLES ANALOG A OUTPUT
Example:
ANALOG OPT STATE OFF
! DISABLES ANALOG B OUTPUT
3.4.2
Setting Options for the Output
The ANALOG function can also configure various aspects of the analog output. The command format is:
Syntax:
ANALOG [STD or OPT] [TOP or BOT] [LG or LN] [a b c d]
ANALOG is the function code for Configure Analog Output
[STD or OPT] specifies the analog output A or B respectively
[TOP or BOT] specifies Channel 1 input or Channel 2 input respectively
[LG or LN] specifies logarithmic (dBm) or linear (Watts) measurement respectively
The command string ends with four numeric variables (with at least one separator character
between each pair of them), which define the relationship between the input power range and
the output voltage range:
a: power level represented by the minimum output voltage,
b: power level represented by the maximum output voltage,
c: minimum output voltage,
d: maximum output voltage
Valid power range numbers are -99.99 to +99.99 [dBm] for LOG, or 0 to 1e3 [Watts] for LIN.
Valid voltage range numbers are 0.00 to +10.00 VDC.
Example:
ANALOG STD BOT LG -80.0, 20.0, 0.0, 10.0
! CONFIGURES THE ANALOG A OUTPUT AS FOLLOWS:
! CHANNEL 2 INPUT, LOGARITHMIC UNITS, -80 TO +20 DBM,
! 0 TO 10 VOLT OUTPUT
Example 2:
ANALOG OPT TOP LN 0.00, 1.00E-3, 0.0, 1.0 ! CONFIGURES THE ANALOG B OUTPUT AS FOLLOWS:
! CHANNEL 1 INPUT, LINEAR UNITS, 0.0 TO 1.0 mW,
! 0 TO 1 VOLT OUTPUT
Содержание 8650B Series
Страница 3: ...Series 8650B Universal Power Meters Publication 31470 001 Rev C November 2 2017 iii ...
Страница 18: ......
Страница 32: ......
Страница 78: ......
Страница 236: ......
Страница 244: ......
Страница 258: ......
Страница 263: ...Menu Structure Publication 31470 001 Rev C November 2 2017 D 3 Figure D 3 Modulation Sensor B Setup Menu Structure ...
Страница 265: ...Menu Stucture Publication 31470 001 Rev C November 2 2017 Page D 5 Figure D 4 Meter Setup Menu Structure ...
Страница 267: ...Publication 31470 001 Rev C November 2 2017 Last page of the document ...