Front Panel Operation
Publication 31470-001, Rev C, November 2, 2017
2-13
Histogram
The Histogram feature displays graphical views of accumulated data distribution over a selective period of time
(See Figure 2-8). From the Power Meter Configuration menu move the cursor to the
Histogram
option and press
[Config]
.
1.
Press
[Select Source]
and select the input channel (8652B Only) or display the line number.
2.
Toggle
Log
or
Lin
with the
[Log/Lin]
softkey.
3.
Press
[Interval]
and select the duration the data is to be acquired. The range is from 1 second to 96 hours, or
continuous.
4.
Press
[OK]
to accept these settings or press
[Cancel]
to cancel the settings, and return to the Power Meter
Configuration menu.
5.
Press
[OK]
from the Power Meter Configuration menu. The Histogram will display full screen as shown in
Figure 2-8: Histogram Illustration
There are additional functions available from the Histogram screen:
1.
Press
[Reset]
to restart sampling.
2.
Press
[Select Mode]
to toggle between distribution display functions.
a.
The Histogram (See Figure 2-8) shows the percentage of time each power level is measured.
b.
The CDF (Cumulative Distribution Function) shows the percentage of time a signal is below a selected
power level (See Figure 2-9). The x-axis displays the power level and the y-axis displays the percentage
of time the power is at or below the power level specified by the x-axis.
c.
The CCDF (Complementary Cumulative Distribution Function) shows the percentage of time a signal is
above a selected power level for more accustomed viewing of a descending slope (See Figure 2-10).
3.
Press the
[Vert/Horiz]
toggle key to select the zoom axis scale, then press
[Zoom In]
or
[Zoom Out]
for the
selected axis.
Содержание 8650B Series
Страница 3: ...Series 8650B Universal Power Meters Publication 31470 001 Rev C November 2 2017 iii ...
Страница 18: ......
Страница 32: ......
Страница 78: ......
Страница 236: ......
Страница 244: ......
Страница 258: ......
Страница 263: ...Menu Structure Publication 31470 001 Rev C November 2 2017 D 3 Figure D 3 Modulation Sensor B Setup Menu Structure ...
Страница 265: ...Menu Stucture Publication 31470 001 Rev C November 2 2017 Page D 5 Figure D 4 Meter Setup Menu Structure ...
Страница 267: ...Publication 31470 001 Rev C November 2 2017 Last page of the document ...