Remote Operation
Publication 31470-001, Rev C, November 2, 2017
3-129
3.19.10
Relative Measurements
In the relative measurement mode, the 8650B saves the current measured power level as a reference.
Subsequent measurements will be expressed relative to this reference level; the measurement units become
dBr (for logarithmic measurement) or % (for linear measurement).
NOTE:
These commands are specific to the active Channel. These commands must be preceded by the
appropriate Channel Configuration using the CH [n] EN command. See Section 3.7.4 for details.
The simple commands associated with relative measurement modes consist of one of three function codes:
Syntax:
[RL0 or RL1 or RL2]
RL0 deactivates the relative measurement mode
RL1 activates the relative mode, and causes the current measured level to be recorded as the
reference level
RL2 activates the relative mode and causes the reference level that was saved under a prior
RL1 command to be used as the reference level. That is, if the relative mode is activated by an
RL1 command, and then deactivated by an RL0 command, the effect of RL2 is to restore the
reference level that was saved in response to the RL1 command
Example:
RL0
! DISABLES THE RELATIVE MODE FOR THE ACTIVE CHANNEL
Example 2:
RL1
! ENABLES THE RELATIVE MODE FOR THE ACTIVE CHANNEL
! SAVES THE CURRENT LEVEL AS A REFERENCE
Example 3:
RL2
! ENABLES THE RELATIVE MODE FOR THE ACTIVE CHANNEL AND
! REPLACE THE CURRENT REFERENCE LEVEL WITH THE PREVIOUS LEVEL
3.19.11
Resolution
NOTE:
These commands are specific to the active Channel. These commands must be preceded by the
appropriate Channel Configuration using the CH [n] EN command. See Section 3.7.4 for details.
Commands which specify measurement resolution are based on the function code RE. The command format is:
Syntax:
RE [a] EN
RE indicates that resolution is being set
[a] indicates the resolution with four values allowed (0, 1, 2 and 3). These specify the number of
digits to the right of the decimal point
EN is a required terminating suffix
Example:
RE 0 EN
! SETS THE DISPLAY RESOLUTION FOR THE ACTIVE CHANNEL TO XX.
Example 2:
RE 1 EN
! SETS THE DISPLAY RESOLUTION FOR THE ACTIVE CHANNEL TO XX.X
Example 3:
RE 2 EN
! SETS THE DISPLAY RESOLUTION FOR THE ACTIVE CHANNEL TO XX.XX
Example 4:
RE 3 EN
! SETS THE DISPLAY RESOLUTION FOR THE ACTIVE CHANNEL TO XX.XXX
Содержание 8650B Series
Страница 3: ...Series 8650B Universal Power Meters Publication 31470 001 Rev C November 2 2017 iii ...
Страница 18: ......
Страница 32: ......
Страница 78: ......
Страница 236: ......
Страница 244: ......
Страница 258: ......
Страница 263: ...Menu Structure Publication 31470 001 Rev C November 2 2017 D 3 Figure D 3 Modulation Sensor B Setup Menu Structure ...
Страница 265: ...Menu Stucture Publication 31470 001 Rev C November 2 2017 Page D 5 Figure D 4 Meter Setup Menu Structure ...
Страница 267: ...Publication 31470 001 Rev C November 2 2017 Last page of the document ...