Introduction
Publication 31470-001, Rev C, November 2, 2017
1-9
800 MHz to 1 GHz Synthesizer (Option 12)
Power Range
+20 to -30 dBm, settable in 1dB steps
Frequency
800 MHz to 1 GHz, settable in 1MHz steps
Power Stability
< 0.1 dB/hour
Frequency Accuracy
±0.05%
NOTE:
Power accuracy for Option 12 is only guaranteed while in calibration mode at 1 GHz, 0 dBm.
Instrumentation Linearity
±0.02 dB over any 20 dB range from -70 to +16 dBm
1
±0.02 dB (±0.05 dB/dB) from +16 to +20 dBm
System Linearity @ 50 MHz
±0.04 dB from -70 to +16 dBm
The following chart shows linearity plus worst case zero set and
noise vs. input power
Instrumentation Linearity (continued)
Temperature Coefficient of Linearity <0.3%/ °C temperature change following Power Sweep
calibration. 24-hour warm-up required
Zeroing Accuracy
Zeroing Set
2,3
<±50 pW with all CW Power Sensors
<±100 pW with Series 804XXA, 806XXA Modulation Power
Sensors
<±200 pW with Series 807XXA Modulation Power Sensors