Front Panel Operation
Publication 31470-001, Rev C, November 2, 2017
2-21
2.3.4.2
Offset
Select this option to enter an offset value in dB when an attenuator or amplifier is installed before the sensor
input. From the Sensor Setup menu:
1.
Press
[Offset]
and enter the value of the offset with the cursor keys.
2.
Press
[OK]
to store the changes, or
[Cancel]
to exit without saving changes.
2.3.4.3
Config
This option is available with Peak and Modulation sensors only.
Peak Sensors
1.
From the Sensor Setup menu select the Peak sensor with the cursor keys.
2.
Press
[Config]
to display the Setup Mode menu. This menu offers two options:
CW
and
Peak
.
3.
Press
[CW]
to select the CW mode for the peak sensor. There is no configuration for peak sensors
operating in the CW mode and the screen will return to the Sensor Setup menu to configure other
sensors.
4.
Press
[Peak]
to select Peak mode and access the peak sensor configuration menu.
5.
Press
[Trig Level]
to configure the trigger level in either the Internal or External Trigger mode.
a.
Internal Trigger
Peak power will be sampled at a point defined by a Trigger Level, a Delay, and a Delay Offset (Refer
to Section 2.4.8.2 for details of these Peak mode parameters). The delay-offset feature is a
convenience in some applications (For example, when measuring pulse width from a point other
than the trigger level, or when comparing the levels of various pulses within a pulse train).
1.)
Press
[Int/Ext Trig]
to select Internal Triggering.
2.)
Set the Trigger Level value with the cursor keys.
3.)
Press
[Delay]
and use the cursor keys to adjust the Trigger Delay (0 to 100 ms). Units will change
automatically.
4.)
Press
[Delay Offset]
and use the cursor keys to adjust the Trigger Delay Offset (0 to 100 ms).
Units will change automatically.
5.)
Press
[OK]
to store the changes, or
[Cancel]
to exit without saving changes.
b.
External Trigger
The External Trigger mode is configured the same as the Internal Trigger mode described above, except
that the Trigger Level is specified in volts rather than dBm.
1.)
Press
[Int/Ext]
Trig to toggle to the External Trigger mode and set the Trigger Level value in volts
with the cursor keys.
2.)
Press
[OK]
to store the changes, or
[Cancel]
to exit without saving changes.
Содержание 8650B Series
Страница 3: ...Series 8650B Universal Power Meters Publication 31470 001 Rev C November 2 2017 iii ...
Страница 18: ......
Страница 32: ......
Страница 78: ......
Страница 236: ......
Страница 244: ......
Страница 258: ......
Страница 263: ...Menu Structure Publication 31470 001 Rev C November 2 2017 D 3 Figure D 3 Modulation Sensor B Setup Menu Structure ...
Страница 265: ...Menu Stucture Publication 31470 001 Rev C November 2 2017 Page D 5 Figure D 4 Meter Setup Menu Structure ...
Страница 267: ...Publication 31470 001 Rev C November 2 2017 Last page of the document ...