Series 8650B Universal Power Meters
3-40
Publication 31470-001, Rev C, November 2, 2017
Syntax:
SENSe<sensor 1 or 2>:TRIGger:LEVel[:MAGnitude]space
< w/INT triggering: -30 to +20 dBm | w/EXT triggering: -0.100 to 5.000 V >
Example:
SENS1:TRIG:LEV 1.7
! SETS PEAK SENSOR 1 TRIGGER LEVEL TO 1.7 VOLTS
! (W/ SENS1:TRIG:SOUR EXT)
Example 2:
SENS1:TRIG:LEV -12.3
! SETS PEAK SENSOR 1 TRIGGER LEVEL TO -12.3 DBM
! (W/ SENS1:TRIG:SOUR INT)
Description:
This command sets the trigger level for the 80350A Series Peak Power Sensors. Under
INTernal triggering the value is in dBm (-30 to +20 dBm, with default of -20 dBm). When
EXTernal triggering is used the value is in Volts (-0.100 to 5.000 V, with a default of 1.700 V).
The trigger level must be lower than the signal that is triggering the peak power sensor. Under
INTernal triggering, a -16 dBm signal must be triggered with a setting that is less than -16 dBm.
EXTernal responds similarly. The input impedance to the EXTernal trigger port on the back of
the sensor housing is not 50
Ω: it is a high impedance to prevent damage from triggering inputs
such as 5.0 V TTL signals. Using a high speed 50
Ω trigger source may cause unwanted signal
reflections and noise on the trigger line. An SMB to SMB attenuator is available from Giga-
tronics to alleviate this condition. The attenuation reduces the reflected noise, and
reduces/eliminates the noisy trigger characteristics.
Be sure to set the trigger level after the INTernal or EXTernal triggering is selected.
Syntax:
SENSe<sensor 1 or 2>:TRIGger:OFFSet[:MAGnitude]space<offset delay in seconds from -50.5e-9 to 104.8e-3>
Example:
SENS1:TRIG:OFFS 0
! SETS PEAK SENSOR 1 TRIGGER OFFSET TIME TO 0 SECONDS
Description:
This command sets the peak power sensor trigger offset time.
Syntax:
SENSe<sensor 1 or 2>:TRIGger:TOTAl[:MAGnitude]?
Example:
SENS1:TRIG:TOTA?
! QUERIES PEAK SENSOR 1 TOTAL TRIGGER DELAY TIME
Response:
Current peak power sensor total trigger delay setting; a single value in seconds.
Description:
This query reports the peak power sensor total trigger delay time. This is equivalent to Delay
plus Delay Offset.
Содержание 8650B Series
Страница 3: ...Series 8650B Universal Power Meters Publication 31470 001 Rev C November 2 2017 iii ...
Страница 18: ......
Страница 32: ......
Страница 78: ......
Страница 236: ......
Страница 244: ......
Страница 258: ......
Страница 263: ...Menu Structure Publication 31470 001 Rev C November 2 2017 D 3 Figure D 3 Modulation Sensor B Setup Menu Structure ...
Страница 265: ...Menu Stucture Publication 31470 001 Rev C November 2 2017 Page D 5 Figure D 4 Meter Setup Menu Structure ...
Страница 267: ...Publication 31470 001 Rev C November 2 2017 Last page of the document ...