Series 8650B Universal Power Meters
1-4
Publication 31470-001, Rev C, November 2, 2017
1.2
Installation
Select the correct operating voltage and install the proper fuse in this housing. Refer to Section 1.2.2, Line
Voltage and Fuse Selection for instructions on how to select the voltage and replace the fuse.
Observe the following Safety Precautions when installing the Series 8650B Universal Power Meter. See Section
1.2.4 for connecting to the rear panel. Also see Section 2.4.3 for instructions on how to connect and calibrate
power sensors.
CAUTION
Do not connect main power to the unit until the required operating voltage and
fuse rating have been checked. The instrument can be damaged if connected to a
source voltage with the line voltage selector set incorrectly.
1.2.1
Safety Precautions
The 8650B has a 3-wire power cord with a 3-terminal polarized plug for connection to the power source and
safety-ground. The ground (or safety ground) is connected directly to the chassis.
WARNING
If a 3-to-2 wire adapter is used, connect the ground lead from the adapter to earth
ground. Failure to do this can cause the instrument to float above earth ground,
posing a shock hazard.
The 8650B is designed for international use with source voltages of 100, 120, 220, or 240 Vac, ±10% at 48 to 440
Hz. The 8650B uses an internationally approved connector that includes voltage selection, fuse, and filter for RFI
protection (See Figure 1-1).
Содержание 8650B Series
Страница 3: ...Series 8650B Universal Power Meters Publication 31470 001 Rev C November 2 2017 iii ...
Страница 18: ......
Страница 32: ......
Страница 78: ......
Страница 236: ......
Страница 244: ......
Страница 258: ......
Страница 263: ...Menu Structure Publication 31470 001 Rev C November 2 2017 D 3 Figure D 3 Modulation Sensor B Setup Menu Structure ...
Страница 265: ...Menu Stucture Publication 31470 001 Rev C November 2 2017 Page D 5 Figure D 4 Meter Setup Menu Structure ...
Страница 267: ...Publication 31470 001 Rev C November 2 2017 Last page of the document ...