Remote Operation
Publication 31470-001, Rev C, November 2, 2017
3-137
Table 3-21: Measurement Error Code Returned in Position AA
Error Code Message
Notes
00
No Error
All OK
01
Cannot zero sensor A
Ensure no RF power to sensor A
02
Cannot zero sensor B
Ensure no RF power to sensor B
03
Sensor A not connected to Calibrator
Connect sensor A to Calibrator
04
Sensor B not connected to Calibrator
Connect sensor B to Calibrator
05
Cannot Cal Sensor A
Check sensor A connection to Calibrator;
reference must be 1.00 mW
06
Cannot Cal sensor B
Check sensor B connection to Calibrator;
reference must be 1.00 mW
21
Over limit
An over-limit condition has occurred
23
Under limit
An under-limit condition has occurred
26
Sensor A unable to synchronize Burst Average Power
(BAP) measurements to a pulse stream
Check measurement setup and RF signal
27
Sensor B unable to synchronize Burst Average Power
(BAP) measurements to a pulse stream
Check measurement setup and RF signal
31
No sensor on Channel A
Connect sensor A, or change channels if B is connected
32
No sensor on Channel B
Connect sensor B, or change channels if A is connected
Содержание 8650B Series
Страница 3: ...Series 8650B Universal Power Meters Publication 31470 001 Rev C November 2 2017 iii ...
Страница 18: ......
Страница 32: ......
Страница 78: ......
Страница 236: ......
Страница 244: ......
Страница 258: ......
Страница 263: ...Menu Structure Publication 31470 001 Rev C November 2 2017 D 3 Figure D 3 Modulation Sensor B Setup Menu Structure ...
Страница 265: ...Menu Stucture Publication 31470 001 Rev C November 2 2017 Page D 5 Figure D 4 Meter Setup Menu Structure ...
Страница 267: ...Publication 31470 001 Rev C November 2 2017 Last page of the document ...