Series 8650B Universal Power Meters
1-2
Publication 31470-001, Rev C, November 2, 2017
1.1.2
Environmental Standards
The Series 8650B instruments are type tested to MIL-PRF-28800F, Class 3 for all departments and agencies of
the Department of Defense applications except as follows:
•
Operating temperature range is 0 °C to 55 °C (Calibrator operating temperature range is 5 °C to 35 °C)
•
Operating the 8651B/8652B Power Meters in a high level RF field (approximately 3 V/m) may degrade
performance, this degradation occurs at measured levels below -36 dBm and when the frequency of
the field is nominally between 50 and 1000 MHz
•
Non-operating (storage) temperature range is -40 °C to +70 °C
•
Relative humidity is limited to 95% non-condensing
1.1.3
Power Requirements
100/120/220/240 Vac ±10%, 48-440 Hz, 20 W, typical. See Section 1.2.1 for details to set the voltage and install
the correct fuse for the area in which the instrument will be used.
1.1.4
Items Furnished
In addition to options and/or accessories specifically ordered, items furnished with the instrument are:
•
Operation Manual (P/N 31470-001)
•
Power Cord (1 ea.)
•
Detachable Sensor Cable (for Model 8651B) (1 ea.), or
•
Detachable Sensor Cables (for Model 8652B) (2 ea.)
1.1.5
Items Required
The 8650B requires an external power sensor; see Appendix B for Power Sensor specifications.
NOTE:
The optional 1 GHz Calibrator is required for operation with Series 807XXA and Series 813XXA
Power Sensors (See Option 12 in Appendix C).
1.1.6
Cooling
No cooling is required if the instrument is operated within its specified operating temperature range (0 to 50 °
C).
1.1.7
Cleaning
The front panel can be cleaned using a cloth dampened with a mild detergent; wipe off the detergent residue
with a damp cloth and dry with a dry cloth. Solvents and abrasive cleaners should not be used.
Содержание 8650B Series
Страница 3: ...Series 8650B Universal Power Meters Publication 31470 001 Rev C November 2 2017 iii ...
Страница 18: ......
Страница 32: ......
Страница 78: ......
Страница 236: ......
Страница 244: ......
Страница 258: ......
Страница 263: ...Menu Structure Publication 31470 001 Rev C November 2 2017 D 3 Figure D 3 Modulation Sensor B Setup Menu Structure ...
Страница 265: ...Menu Stucture Publication 31470 001 Rev C November 2 2017 Page D 5 Figure D 4 Meter Setup Menu Structure ...
Страница 267: ...Publication 31470 001 Rev C November 2 2017 Last page of the document ...