Series 8650B Universal Power Meters
3-104
Publication 31470-001, Rev C, November 2, 2017
3.15
Limits
NOTE:
Both limits need to be set and LL must always be lower than LH. The reason is that the default
settings are 0 dBm for LL and LH.
NOTE:
These commands are specific to a predefined Channel. These commands must be preceded by the
appropriate Channel Configuration using the CH [n] EN command. See Section 3.7.4 for details.
3.15.1
Setting Limits
Commands which set limits are based on the LH and LL function codes. The command format is:
Syntax:
[LH or LL] [
n
] EN
LH specifies the high limit.
LL specifies the low limit.
[
n
] is a limit value, expressed in dBm or dB as appropriate.
EN is a required terminating suffix.
Example:
LH 12.34 EN
! SETS HIGH LIMIT TO +12.34 DB
Example 2:
LL -2.58 EN
! SETS LOW LIMIT TO -2.58 DB
3.15.2
Activating Limits
Limit-checking is activated or deactivated by simple commands consisting of one of two function codes:
Syntax:
[LM0 or LM1]
[LM0] disables limit checking
[LM1] enables limit checking
Example:
LM0
! DISABLES LIMIT CHECKING
Example 2:
LM1
! ENABLES LIMIT CHECKING
Before enabling limit checking (LM1), set the high and low limits (LH and LL). Once enabled, the Over/Under
Limit bit of the Status Byte Register (Bit 7) will signal a limit violation condition (see Section 3.19.13.1). In
addition, the Status Message AA bytes will indicate the occurance of an Over-Limit condition (Measurement
Error Code 21), or an Under-Limit condition (Measurement Error Code 23). The Status Message byte L also
contains the current limit status condition. 0 indicates within limits, 1 indicates over high limit, and 2 indicates
under low limit. See Section 3.19.13.3 for more information.
The LCD display will indicate a too high condition with Over Limit displayed to the right of the reading, and
Under Limit displayed to the right of the reading for a too low condition.
Содержание 8650B Series
Страница 3: ...Series 8650B Universal Power Meters Publication 31470 001 Rev C November 2 2017 iii ...
Страница 18: ......
Страница 32: ......
Страница 78: ......
Страница 236: ......
Страница 244: ......
Страница 258: ......
Страница 263: ...Menu Structure Publication 31470 001 Rev C November 2 2017 D 3 Figure D 3 Modulation Sensor B Setup Menu Structure ...
Страница 265: ...Menu Stucture Publication 31470 001 Rev C November 2 2017 Page D 5 Figure D 4 Meter Setup Menu Structure ...
Страница 267: ...Publication 31470 001 Rev C November 2 2017 Last page of the document ...