Series 8650B Universal Power Meters
3-106
Publication 31470-001, Rev C, November 2, 2017
3.16
Measurement Collection Modes (Standard)
3.16.1
Measurement Triggering
Trigger modes determine when a measurement will be made. Four simple commands consisting of one of four
function codes select the desired mode:
Syntax:
[TR0 or TR1 or TR2 or TR3]
All four modes discussed here are standard measurement collection modes (as opposed to the GPIB fast modes
described in Section 3.17), and use the standard data output format.
3.16.1.1
Trigger Hold (TR0)
This command places the instrument in standby mode. The LCD display is frozen at the current values. The
display will be updated when the instrument receives a TR1 or TR2 command. To resume the normal free run
mode of the instrument and display, use the TR3 command. During the standby mode, the instrument continues
to make measurements and update the internal digital filter, but does not update the display or the GPIB buffer.
Example:
TR0
! SELECTS THE TRIGGER HOLD MODE
3.16.1.2
Trigger Immediate (TR1)
This command triggers a single reading; the reading is added to the internal digital filter. An ENTER statement
will return the updated filter power level. After a TR1 command, the instrument returns to the standby mode.
Example:
TR1
! TRIGGER A SINGLE MEASUREMENT
3.16.1.3
Trigger Immediate with Full Averaging (TR2)
This mode triggers a new series of readings; enough to update the digital filter for a noise free reading at the
current power level. An ENTER statement will return the fully updated filter power level. After a TR2 command,
the instrument returns to the standby mode.
Example:
TR2
! TRIGGER A FULL MEASUREMENT, WITH SETTLING
3.16.1.4
Free Run (TR3)
This free run trigger mode (which is the default mode) allows the user to read the power at any time with an
ENTER statement. There is no need to send the TR3 command again. Multiple ENTER statements can be
executed. The power meter will return the present power level just as if looking at the LCD display.
Example:
TR3
! FREE RUN TRIGGER MODE
Содержание 8650B Series
Страница 3: ...Series 8650B Universal Power Meters Publication 31470 001 Rev C November 2 2017 iii ...
Страница 18: ......
Страница 32: ......
Страница 78: ......
Страница 236: ......
Страница 244: ......
Страница 258: ......
Страница 263: ...Menu Structure Publication 31470 001 Rev C November 2 2017 D 3 Figure D 3 Modulation Sensor B Setup Menu Structure ...
Страница 265: ...Menu Stucture Publication 31470 001 Rev C November 2 2017 Page D 5 Figure D 4 Meter Setup Menu Structure ...
Страница 267: ...Publication 31470 001 Rev C November 2 2017 Last page of the document ...