Remote Operation
Publication 31470-001, Rev C, November 2, 2017
3-99
3.12.2
Frequency
Cal factors are stored in the sensor’s EEPROM by frequency. Specifying a frequency causes the 8650B to apply
the cal factor appropriate to that frequency. To cancel the use of cal factors, specify the calibration frequency of
the selected sensor (this is the frequency of the front panel Calibrator reference output, and has a cal factor of
zero). For the 81305A and the 80701A Series sensors this frequency is 1 GHz. For all other power sensors this
frequency is 50 MHz.
Commands which specify a frequency are based on the FR function. The command format is:
Syntax:
[AE or BE] FR [n] [HZ or KZ or MZ or GZ]
[AE or BE] prefix specifies Sensor A or Sensor B
FR specifies a frequency value
[n] specifies the frequency value (the units are Hz, kHz, MHz, or GHz, depending on the
terminating suffix used)
[HZ or KZ or MZ or GZ] is a required terminating suffix
Example:
AE FR 5.67 GZ
! FREQUENCY FOR SENSOR A IS 5.67 GHZ
Example 2:
AE FR 1.0E9 HZ
! FREQUENCY FOR SENSOR A IS 1.0 GHZ
Example 3:
BE FR 84.6 MZ
! FREQUENCY FOR SENSOR B IS 84.6 MHZ
Содержание 8650B Series
Страница 3: ...Series 8650B Universal Power Meters Publication 31470 001 Rev C November 2 2017 iii ...
Страница 18: ......
Страница 32: ......
Страница 78: ......
Страница 236: ......
Страница 244: ......
Страница 258: ......
Страница 263: ...Menu Structure Publication 31470 001 Rev C November 2 2017 D 3 Figure D 3 Modulation Sensor B Setup Menu Structure ...
Страница 265: ...Menu Stucture Publication 31470 001 Rev C November 2 2017 Page D 5 Figure D 4 Meter Setup Menu Structure ...
Страница 267: ...Publication 31470 001 Rev C November 2 2017 Last page of the document ...